Accuracy of Digora System in Detecting Artificial Peri-implant Bone Defects
Authors:
- Yukiko MatsudaYMYukiko Matsuda
- Tomomi HanazawaTHTomomi Hanazawa
- Kenji SekiKSKenji Seki
- Tsukasa SanoTSTsukasa Sano
- Masahiko OzekiMOMasahiko Ozeki
- Tomohiro OkanoTOTomohiro Okano
- 1 December 2001
- journal article
- research article
- Published by Wolters Kluwer Health in Implant Dentistry
- Vol. 10 (4) , 265-271
Abstract
This study was done to compare the diagnostic accuracy in detecting simulated intrabony defects around fixtures using Digora (DIG; Sordex Orion Corporation, Helsinki, Finland) compared with Ektaspeed Plus film (PLS; Eastman Kodak Co., Rochester, NY). Three titanium implant fixtures were placed in molar areas of three cadavers. Bone defects were created in the interproximal alveolar crest. Exposure time was adjusted to a PLS film and reduced to 1/5 only for DIG (1/5 DIG). The results of four observers were assessed. Sensitivity, specificity, accuracy, and under the receiver operator char-acteristic curve (ROC) area were evaluated. Statistical analyses were performed by using Friedman test and one-way ANOVA test. Mean sensitivity/specificity were 0.60/0.85 (DIG), 0.54/0.81 (1/5 DIG), and 0.64/0.58 (PLS). There were no statistically significant differences in the diagnostic accuracies. Digora had an equivalent performance to radiographic film in detecting intrabony defects adjacent to the implants, notwithstanding the amount of 1/5 of the exposure time.Keywords
This publication has 19 references indexed in Scilit:
- Effects of Guided Bone Regeneration Around Commercially Pure Titanium and Hydroxyapatite‐Coated Dental Implants. I. Radiographic AnalysisJournal of Periodontology, 1997
- On the dynamic range of different X-ray photon detectors in intra-oral radiography. A comparison of image quality in film, charge-coupled device and storage phosphor systems.Dentomaxillofacial Radiology, 1996
- Accuracy of caries diagnosis in digital images from charge-coupled device and storage phosphor systems: an in vitro study.Dentomaxillofacial Radiology, 1995
- Future trends in dental radiologyOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 1995
- Radiological evaluation of marginal bone loss at tooth surfaces facing single Brånemark implantsClinical Oral Implants Research, 1993
- Exposure reduction in cephalography with a digital photostimulable phosphor imaging system.Dentomaxillofacial Radiology, 1993
- A comparative study of the clinical efficacy of Screw Vent irnnlants versus Brånemark fixtures, installed in a periodontal clinicClinical Oral Implants Research, 1992
- The Excessive Loss of Branemark Fixtures in Type IV Bone: A 5‐Year AnalysisJournal of Periodontology, 1991
- Computed panoramic tomography with scanning laser-stimulated luminescenceOral Surgery, Oral Medicine, Oral Pathology, 1985
- A 15-year study of osseointegrated implants in the treatment of the edentulous jawInternational Journal of Oral Surgery, 1981