Structural characterization and in-vivo reliability evaluation of silicon microneedles
Author:
- 2 February 2014
- journal article
- research article
- Published by Springer Nature in Biomedical Microdevices
- Vol. 16 (3) , 333-343
Abstract
This work presents an analysis of the failure mechanisms, structural properties and reliability of wet-etched silicon microneedles, which have wide-ranging applications in transdermal delivery, sensing and diagnostics. For the first time, in-vivo skin insertion forces are measured and the structural properties of individual silicon microneedles are assessed using both compression and shear tests. Compressive failure of this particular microneedle design does not occur because of buckling, but instead is predominantly due to progressive fracture along the relatively weak {111} crystal plane. Compressive and shear failure strengths are experimentally determined to be (2.9 +/− 0.3) GPa and (9.2 +/− 0.2) MPa, respectively. It is also shown that basic mechanical tests that are commonly used in the field of microneedle development may significantly underestimate safety factors for this type of needle due to the unrepresentative nature of the interaction of a rigid surface with the needle tips. Therefore, a new figure-of-merit for the reliability of such microneedles is proposed, which is based on the ratio of material failure strength to peak stress during skin insertion. The distribution of forces over the sharp, conical needle tip during skin penetration leads to a very large safety margin of over 700, and a correspondingly high degree of reliability when applied to in-vivo human tissue.Keywords
This publication has 34 references indexed in Scilit:
- Dual drug-eluting stentJournal of Controlled Release, 2012
- Determination of parameters for successful spray coating of silicon microneedle arraysInternational Journal of Pharmaceutics, 2011
- Microneedle Array Design Determines the Induction of Protective Memory CD8+ T Cell Responses Induced by a Recombinant Live Malaria Vaccine in MicePLOS ONE, 2011
- Dissolving polymer microneedle patches for influenza vaccinationNature Medicine, 2010
- Axial and shear fracture strength evaluation of silicon microneedlesMicrosystem Technologies, 2010
- Microneedle-based drug delivery systems: Microfabrication, drug delivery, and safetyDrug Delivery, 2010
- Autonomous capillary system for one-step immunoassaysBiomedical Microdevices, 2008
- The evolution from convex corner undercut towards microneedle formation: theory and experimental verificationJournal of Micromechanics and Microengineering, 2006
- Process optimization and characterization of silicon microneedles fabricated by wet etch technologyMicroelectronics Journal, 2005
- Analytic solutions for tapered column bucklingComputers & Structures, 1988