Enhanced ventricular tachycardia substrate resolution with a novel omnipolar high-density mapping catheter: the omnimapping study

Authors:

Abstract
Defining diastolic slow-conduction channels within the borderzone (BZ) of scar-dependent re-entrant ventricular tachycardia (VT) is key for effective mapping and ablation strategies. Understanding wavefront propagation is driving advances in high-density (HD) mapping. The newly developed Advisor™ HD Grid Mapping Catheter (HD GRID) has equidistant spacing of 16, 1 mm electrodes in a 4 × 4 3 mm interspaced arrangement allowing bipolar recordings along and uniquely across the splines (orthogonal vector) to facilitate substrate mapping in a WAVE configuration (WAVE). The purpose of this study was to determine the relative importance of the WAVE configuration compared to the STANDARD linear-only bipolar configuration (STANDARD) in defining VT substrate.

This publication has 18 references indexed in Scilit: