Distal Locking Using an Electromagnetic Field–Guided Computer-Based Real-Time System for Orthopaedic Trauma Patients
Authors:
- Jason J. HalvorsonJHJason J. Halvorson
- Aaron T. ScottASAaron T. Scott
- Beth P. SmithBSBeth P. Smith
- Gregory B. RussellGRGregory B. Russell
- Riyaz H. JinnahRJRiyaz H. Jinnah
- Anna N. MillerAMAnna N. Miller
- Eben A. CarrollECEben A. Carroll
- 1 July 2013
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Orthopaedic Trauma
- Vol. 27 (7) , 367-372
Abstract
To compare the efficacy of distal interlocking during intramedullary nailing using a freehand technique versus an electromagnetic field real-time system (EFRTS). A prospective, randomized controlled trial. Level I academic trauma center. Patients older than 18 years who sustained a femoral or tibial shaft fracture amenable to antegrade intramedullary nailing were prospectively enrolled between August 2010 and November 2011. Exclusion criteria included injuries requiring retrograde nailing and open wounds near the location of the distal interlocks (distal third of the femur, knee, or distal tibia). Each patient had 2 distal interlocking screws placed: one using the freehand method and the other using EFRTS. Techniques were compared on procedural time and number of interlocking screw misses. Two time points were measured: time 1 (time to find perfect circles/time from wand placement to drill initiation) and time 2 (drill initiation until completion of interlocking placement). Twenty-four tibia and 24 femur fractures were studied. EFRTS proved faster at times 1 and 2 (P < 0.0001 and P < 0.0002) and total time (P < 0.0001). This difference was larger for junior residents, though reached statistical significance for senior residents. Senior residents were faster with the freehand technique compared with junior residents (P < 0.004), but the 2 were similar using EFRTS (P = 0.41). The number of misses was higher with free hand compared with EFRTS (P = 0.02). These results suggest that EFRTS is faster than the traditional freehand technique and results in fewer screw misses. Therapeutic Level II. See Instructions for Authors for a complete description of levels of evidence.Keywords
This publication has 19 references indexed in Scilit:
- Intramedullary Nailing of Femoral Shaft Fractures: Current ConceptsJournal of the American Academy of Orthopaedic Surgeons, 2009
- Efficacy of an aiming device for the placement of distal interlocking screws in trochanteric fixation nailingArchives of orthopaedic and trauma surgery, 2008
- Nail over nail technique for distal locking of femoral intramedullary nailsInternational Orthopaedics, 2008
- Targeting device for intramedullary nails: A new high-stable mechanical guideInjury, 2008
- Distal Locking of Tibial NailsClinical Orthopaedics and Related Research, 2008
- The effectiveness of the external distal aiming device in intramedullary fixation of tibial shaft fracturesArchives of orthopaedic and trauma surgery, 2007
- Tibial intramedullary nail distal interlocking screw placement: comparison of the free-hand versus distally-based targeting device techniquesInjury, 2001
- Novel Computer-Assisted Fluoroscopy System for Intraoperative Guidance: Feasibility Study for Distal Locking of Femoral NailsJournal of Orthopaedic Trauma, 2001
- Reamed Intramedullary Nailing of the FemurThe Journal of Trauma: Injury, Infection, and Critical Care, 1999
- A new mechanical aiming device for the placement of distal interlocking screws in femoral nailsArchives of orthopaedic and trauma surgery, 1998