Using spectral reflectance to distinguish between tracheal and oesophageal tissue: applications for airway management
Authors:
- 21 January 2019
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 74 (3) , 340-347
Abstract
Summary: Proper placement of the tracheal tube requires confirmation, and the predominant method in addition to clinical signs is the presence of end‐tidal carbon dioxide. Such is the importance of confirmation that novel methods may also have a place. We previously demonstrated using ex‐vivo swine tissue a unique spectral reflectance characteristic of tracheal tissue that differs from oesophageal tissue. We hypothesised that this characteristic would be present in living swine tissue and human cadavers. Reflectance spectra in the range 500–650 nm were captured using a customised fibreoptic probe, compact spectrometer and white light source from both the trachea and the oesophagus in anesthetised living swine and in human cadavers. A tracheal detection algorithm using ratio comparisons of reflectance was developed. The existence of the unique tracheal characteristic in both in‐vivo swine and cadaver models was confirmed (p < 0.0001 for all comparisons between tracheal and oesophageal tissue at all target wavelengths in both species). Furthermore, our proposed tracheal detection algorithm exhibited a 100% positive predictive value in both models. This has potential utility for incorporation into airway management devices.Keywords
Funding Information
- Medical Research and Materiel Command
- Combat Casualty Care Research Program
This publication has 27 references indexed in Scilit:
- Global Capnography Project (GCAP): implementation of capnography in Malawi – an international anaesthesia quality improvement projectAnaesthesia, 2018
- Strategies for the prevention of airway complications – a narrative reviewAnaesthesia, 2017
- Capnography during cardiopulmonary resuscitation: Current evidence and future directionsJournal of Emergencies, Trauma, and Shock, 2014
- A prospective study of physician pre-hospital anaesthesia in trauma patients: oesophageal intubation, gross airway contamination and the ‘quick look’ airway assessmentBMC Anesthesiology, 2013
- The Importance of First Pass Success When Performing Orotracheal Intubation in the Emergency DepartmentAcademic Emergency Medicine, 2013
- Association Between Repeated Intubation Attempts and Adverse Events in Emergency Departments: An Analysis of a Multicenter Prospective Observational StudyAnnals of Emergency Medicine, 2012
- Prehospital Intubations and Mortality: A Level 1 Trauma Center PerspectiveAnesthesia & Analgesia, 2009
- Prehospital Standardization of Medical Airway Management: Incidence and Risk Factors of Difficult AirwayAcademic Emergency Medicine, 2006
- The Effectiveness of Out-of-Hospital Use of Continuous End-Tidal Carbon Dioxide Monitoring on the Rate of Unrecognized Misplaced Intubation Within a Regional Emergency Medical Services SystemAnnals of Emergency Medicine, 2005
- Rapid detection of oesophageal intubation: take care when using colorimetric capnometryActa Anaesthesiologica Scandinavica, 2002