In the present paper, the preparation and characterization of an improved solid-state pH sensor are described. The sensor is based on anodically electrodeposited iridium oxide film, as a pH-sensing layer. Merits of the present sensor include (i) excellent adhesion of the pH sensitive layer to the substrate, (ii) excellent reproducibility of sensor fabrication, (iii) faster preparation procedure, and (iv) low cost of the titanium substrate. These advantages are realized by combining acid-etched titanium as the electrode substrate with an optimized electrodeposition solution consisting of IrCl4 as an iridium source, hydrogen peroxide, potassium oxalate, and potassium carbonate. Heating the electrodeposition solution to 90 °C reduced the time required for solution development from ∼3 days to 10 min. The pH-sensing layer is protected with a layer of Nafion and a microporous polyester membrane. The improved sensor showed a super-Nernstian response (−73.7 ± 1.2 mV/pH unit) in the pH range of 1.5−11.5. The present pH sensor, fabricated in a tubular form, is used as a detector in a flow injection analysis (FIA) system for pH measurements. Optimization of the FIA experimental parameters resulted in a linear dependence of peak heights on the pH of the injected samples in the pH range of 2−11.Keywords
This publication has 16 references indexed in Scilit:
- Direct measurement of ph within marine biofilms on passive metalsBiofouling, 2000
- An on-line pH monitoring system for flow solution analysis with pH-ISFETs and flow-through cellSensors and Actuators B: Chemical, 2000
- An integrated module for sensing pO2, pCO2, and pHAnalytica Chimica Acta, 2000
- A solid-state pH sensor based on a Nafion-coated iridium oxide indicator electrode and a polymer-based silver chloride reference electrodeSensors and Actuators B: Chemical, 1994
- Measurement of pH by flow injection over a wide dynamic range with PVC/neutral carrier electrodesAnalytica Chimica Acta, 1989
- Flow-through pH-ISFET detector for flow-injection analysisAnalytica Chimica Acta, 1989
- Contents pagesThe Analyst, 1988
- Continuous-flow real-time and flow injection determination of rainwater pH with a poly(vinyl chloride)-based tubular membrane electrodeAnalytical Chemistry, 1987
- Electrodes in series. Simultaneous flow injection determination of chloride and pH with ion-selective electrodesThe Analyst, 1986
- Evaluation of critical parameters for measurement of pH by flow injection analysis determination of pH in soil extractsAnalytica Chimica Acta, 1985