Cardiac output measured by transthoracic echocardiography and Swan-Ganz catheter. A comparative study in mechanically ventilated patients with high positive end-expiratory pressure
Authors:
Open Access
- 1 January 2019
- journal article
- Published by Associação de Medicina Intensiva Brasileira in Revista Brasileira de Terapia Intensiva
- Vol. 31 (4) , 474-482
Abstract
To compare cardiac output measurements by transthoracic echocardiography and a pulmonary artery catheter in mechanically ventilated patients with high positive end-expiratory pressure. To evaluate the effect of tricuspid regurgitation. Sixteen mechanically ventilated patients were studied. Cardiac output was measured by pulmonary artery catheterization and transthoracic echocardiography. Measurements were performed at different levels of positive end-expiratory pressure (10cmH2O, 15cmH2O, and 20cmH2O). The effect of tricuspid regurgitation on cardiac output measurement was evaluated. The intraclass correlation coefficient was studied; the mean error and limits of agreement were studied with the Bland-Altman plot. The error rate was calculated. Forty-four pairs of cardiac output measurements were obtained. An intraclass correlation coefficient of 0.908 was found (p < 0.001). The mean error was 0.44L/min for cardiac output values between 5 and 13L/min. The limits of agreement were 3.25L/min and -2.37L/min. With tricuspid insufficiency, the intraclass correlation coefficient was 0.791, and without tricuspid insufficiency, 0.935. Tricuspid insufficiency increased the error rate from 32% to 52%. In patients with high positive end-expiratory pressure, cardiac output measurement by transthoracic echocardiography is comparable to that with a pulmonary artery catheter. Tricuspid regurgitation influences the intraclass correlation coefficient. In patients with high positive end-expiratory pressure, the use of transthoracic echocardiography to measure cardiac output is comparable to invasive measures.Keywords
This publication has 51 references indexed in Scilit:
- Left Ventricular Systolic and Diastolic Function in Obstructive Sleep ApneaCirculation: Heart Failure, 2012
- The Effect of Pulmonary Artery Catheter Use on Costs and Long-Term Outcomes of Acute Lung InjuryPLOS ONE, 2011
- Determination of the Precision Error of the Pulmonary Artery Thermodilution Catheter Using an In Vitro Continuous Flow Test RigAnesthesia & Analgesia, 2011
- Respiratory and haemodynamic changes during decremental open lung positive end-expiratory pressure titration in patients with acute respiratory distress syndromeCritical Care, 2009
- Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac outputCritical Care, 2009
- Comparison of Doppler Echocardiography With Thermodilution for Assessing Cardiac Output in Advanced Congestive Heart FailureThe American Journal of Cardiology, 1996
- Determination of cardiac output using acoustic quantification in critically III patientsThe American Journal of Cardiology, 1994
- Echocardiographic analysis of cardiac function during high PEEP ventilationIntensive Care Medicine, 1994
- Errors in the measurement of cardiac output by thermodilutionCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1993
- The effect of variations of pulsed Doppler sampling site on calculation of cardiac output: an experimental study in open-chest dogs.Circulation, 1983