Comparison of modeling accuracy between Radixact ® and CyberKnife ® Synchrony ® respiratory tracking system
Authors:
- 2 September 2021
- journal article
- research article
- Published by IOP Publishing in Biomedical Physics & Engineering Express
- Vol. 7 (6) , 067001
Abstract
Synchrony Respiratory Tracking system adapted from CyberKnife has been introduced in Radixact to compensate the tumor motion caused by respiration. This study aims to compare the modeling accuracy of the Synchrony system between Radixact and CyberKnife. Two Synchrony plans based on fiducial phantoms were created for CyberKnife and Radixact, respectively. Different respiratory motion traces were used to drive a motion platform to move along the superoinferior and left-right direction. The cycle time and the amplitude of target/surrogate motion of one selected motion trace were scaled to investigate the dependence of modeling accuracy on the motion characteristic. The predicted target position, the correlation error, potential difference (Radixact only) and standard error (CyberKnife only) were extracted from raw data or log files of the two systems. The modeling accuracy was evaluated by calculating the root-mean-square (RMS) error between the predicted target positions and the input motion trace. A threshold T95 within which 95% of the potential difference or the standard error lay was defined and evaluated. Except for the motion trace with a small amplitude and a good (linear) correlation between target and surrogate motion, Radixact showed smaller RMS errors than CyberKnife. The RMS error of both systems increased with the motion amplitude and showed a decreasing trend with the increasing cycle time. No correlation was found between the RMS error and the amplitude of surrogate motion. T95 could be a good estimator of modeling accuracy for CyberKnife rather than Radixact. The correlation error defined in Radixact were largely affected by the number of fiducial markers and the setup error. In general, the modeling accuracy of the Radixact Synchrony system is better than that of the CyberKnife Synchrony system under unfavorable conditions.Keywords
This publication has 26 references indexed in Scilit:
- Verification of Accuracy of CyberKnife Tumor-tracking Radiation Therapy Using Patient-specific Lung PhantomsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2015
- Stereotactic radiotherapy (SABR) for the treatment of primary non-small cell lung cancer; Systematic review and comparison with a surgical cohortRadiotherapy and Oncology, 2013
- Correlation and prediction uncertainties in the CyberKnife Synchrony respiratory tracking systemMedical Physics, 2011
- CyberKnife for hilar lung tumors: report of clinical response and toxicityJournal of Hematology & Oncology, 2010
- IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119Medical Physics, 2009
- Stereotactic radiotherapy with real-time tumor tracking for non-small cell lung cancer: Clinical outcomeRadiotherapy and Oncology, 2009
- An analysis of thoracic and abdominal tumour motion for stereotactic body radiotherapy patientsPhysics in Medicine & Biology, 2008
- Image-guided Hypo-fractionated Stereotactic Radiosurgery to Spinal LesionsNeurosurgery, 2001
- TomotherapySeminars in Radiation Oncology, 1999
- Tomotherapy: A new concept for the delivery of dynamic conformal radiotherapyMedical Physics, 1993