Experimental assessment of elemental analyzer isotope ratio mass spectrometry normalization methodologies for environmental stable isotopes
- 24 June 2024
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 38 (17) , e9837
Abstract
Rationale: In stable isotope mass spectrometry, isotope delta values are normalized to internationally recognized reference scales using a combination of certified and in‐house isotope reference materials. Numerous techniques exist for performing this normalization, but these methodologies need to be experimentally assessed to compare their effect on reproducibility of isotope results. Methods: We tested normalization methods by the number of reference materials used, their matrix, their isotope range, and whether normalization required extrapolating beyond the isotope range. We analyzed eight commercially available isotope reference materials on a ThermoFinnigan Delta‐V isotope ratio mass spectrometer (IRMS) and an Elementar VisION IRMS for nitrogen and carbon isotope composition via solid combustion with an elemental analyzer and computed every possible isotope normalization (n = 612). Additionally, we assessed how sample matrix affected linearity effects on both instruments using five in‐house reference materials. Results: Normalizations exhibited the best performance when the reference materials spanning an isotope range of at least 20‰ were matrix matched with the samples and did not require extrapolation beyond the calibration curve. When these conditions were not met, the number of reference materials used had a significant effect on accuracy, with normalizations composed of two reference materials exhibiting particularly inconsistent performance at isotope ranges below 20‰. Linearity effects were found to exceed instrument precision by two orders of magnitude irrespective of matrix type and were not predicted by working gas diagnostics. Conclusions: Interlaboratory comparability of isotope results is improved when operators of elemental analyzer isotope ratio mass spectrometry (EAIRMS) systems select reference materials spanning an isotope range of at least 20‰. Additionally, using three or more isotopic reference materials, avoiding extrapolation beyond the range of the normalization curve, and matching the matrix of the reference materials to the samples improve normalizations.Keywords
Funding Information
- U.S. Environmental Protection Agency
- Oak Ridge Institute for Science and Education
This publication has 33 references indexed in Scilit:
- Calibration hierarchies for light element isotope delta reference materialsRapid Communications in Mass Spectrometry, 2019
- Practical and theoretical considerations for the determination of δ 13 C VPDB values of methylmercury in the environmentRapid Communications in Mass Spectrometry, 2019
- Guidelines and recommended terms for expression of stable‐isotope‐ratio and gas‐ratio measurement resultsRapid Communications in Mass Spectrometry, 2011
- Error propagation in normalization of stable isotope data: a Monte Carlo analysisRapid Communications in Mass Spectrometry, 2010
- Stable Isotope Forensics2010
- Normalization of measured stable isotopic compositions to isotope reference scales – a reviewRapid Communications in Mass Spectrometry, 2007
- After two decades a second anchor for the VPDB δ 13 C scaleRapid Communications in Mass Spectrometry, 2006
- Two new organic reference materials for δ 13 C and δ 15 N measurements and a new value for the δ 13 C of NBS 22 oilRapid Communications in Mass Spectrometry, 2003
- Referencing strategies and techniques in stable isotope ratio analysisRapid Communications in Mass Spectrometry, 2001
- Reference materials by isotope ratio mass spectrometryMass Spectrometry Reviews, 1993