This paper describes a method for automatic optimisation of parameters (AOP) in digital mammography systems. Using a model of the image chain, contrast to noise ratio (CNR) and average glandular dose (AGD) are computed for possible X-ray parameters and breast types. The optimisation process consists of the determination of the operating points providing the lowest possible AGD for each CNR level and breast type. The proposed metric for the dose used in the design of an AOP mode is the resulting dose to the population, computed by averaging the AGD values over the distribution of breast types in the population. This method has been applied to the automatic exposure control of new digital mammography equipment. Breast thickness and composition are estimated from a low dose pre-exposure and used to index tables containing sets of optimised operating points. The resulting average dose to the population ranges from a level comparable to state-of-the-art screen/film mammography to a reduction by a factor of two. Using this method, both CNR and dose are kept under control for all breast types, taking into consideration both individual and collective risk.Keywords
This publication has 10 references indexed in Scilit:
- Experimental investigation of the dose and image quality characteristics of a digital mammography imaging systemMedical Physics, 2003
- Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study.British Journal of Radiology, 2000
- Full breast digital mammography with an amorphous silicon‐based flat panel detector: Physical characteristics of a clinical prototypeMedical Physics, 2000
- Experimental and theoretical spectral optimization for digital mammography1999
- Multiparameter optimization of mammography with alternative x-ray sources1995
- Normalized average glandular dose in molybdenum target-rhodium filter and rhodium target-rhodium filter mammography.Radiology, 1994
- Spectral dependence of glandular tissue dose in screen-film mammography.Radiology, 1991
- An approach to specifying a minimum dose system for mammography using multiparameter optimization techniquesMedical Physics, 1985
- The Monte Carlo calculation of integral radiation dose in xeromammographyPhysics in Medicine & Biology, 1980
- Absorbed Radiation Dose in MammographyRadiology, 1979