Broadband optical ultrasound sensor with a unique open-cavity structure
Authors:
Open Access
- 1 January 2011
- journal article
- research article
- Published by SPIE-Intl Soc Optical Eng in Journal of Biomedical Optics
- Vol. 16 (1) , 017001-6
Abstract
There is a pervasive demand for high-resolution imaging techniques for comprehensive morphological and functional imaging of critical biological organs covered by thick tissue layers. Optical microscopy normally provides superior resolution but with shallow imaging depth in scattering media. In contrast, acoustic imaging enables deep tissue imaging but with relatively low spatial resolution, restricted by the frequency range employed in conventional diagnostic ultrasound imaging. In principle, ultrasound imaging resolution can be improved by using higher frequencies (e.g., 50 MHz). However, for a conventional piezoelectric-based ultrasound sensor, higher operation frequency demands smaller effective sensor size,1 which imposes several technological difficulties in the fabrication of miniature piezoelectric ultrasound sensors. The challenges include dicing piezoceramics to micrometer-size elements,2 making electrical connections to the small elements, cross talk between elements, and increased detection noise due to reduced element size.3 Although recently developed miniature polyvinylidene fluoride (PVDF) membrane hydrophones can potentially achieve 4, 5 a frequency bandwidth beyond 140 MHz, demanding electrical connections and cross talk persist if these membrane hydrophones are intended for imaging arrays. Furthermore, it has been reported that the noise-equivalent pressure (NEP) of a 50-μm-diam Fabry–Pérot optical ultrasound sensor is comparable to that of 1-mm-diam PVDF membrane hydrophone.6 Given that the sensitivity of a membrane hydrophone decreases as its area becomes smaller due to lower capacitance,7 the Fabry–Pérot ultrasound sensor can potentially outperform the PVDF membrane sensor with the same active area.8Keywords
Funding Information
- U.S. National Institutes of Health (R21EB008765)
This publication has 22 references indexed in Scilit:
- Sensitive molecular binding assay using a photonic crystal structure in total internal reflectionOptics Express, 2008
- Atherosclerotic plaque characterization with optoacoustic imaging2005
- Optoacoustic imaging using thin polymer étalonApplied Physics Letters, 2005
- Yield strength of thin-film parylene-CMicrosystem Technologies, 2004
- Ultrahigh-sensitivity wideband Fabry-Perot ultrasound sensors as an alternative to piezoelectric PVDF transducers for biomedical photoacoustic detection2004
- Fabry Perot polymer film fibre-optic hydrophones and arrays for ultrasound field characterisationJournal of Physics: Conference Series, 2004
- Microfabrication of thick and bulk PZT materials for piezoelectric actuator2002
- Coronary Plaque Classification With Intravascular Ultrasound Radiofrequency Data AnalysisCirculation, 2002
- Miniature optical fibre ultrasonic hydrophone usinga Fabry-Perot polymer film interferometerElectronics Letters, 1997
- Are hydrophones of diameter 0.5 mm small enough to characterise diagnostic ultrasound equipment?Physics in Medicine & Biology, 1989