Comparative Analysis of the Visual Performance After Cataract Surgery With Implantation of a Bifocal or Trifocal Diffractive IOL
Authors:
- Peter MojzisPMPeter Mojzis
- Lucia KukuckovaLKLucia Kukuckova
- Katarina MajerovaKMKatarina Majerova
- Katerina LiehneovaKLKaterina Liehneova
- David P. PiñeroDavid P. Piñero
- 1 October 2014
- journal article
- research article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 30 (10) , 666-672
Abstract
PURPOSE: To evaluate and compare the visual, refractive, contrast sensitivity, and aberrometric outcomes with a diffractive bifocal and trifocal intraocular lens (IOL) of the same material and haptic design. METHODS: Sixty eyes of 30 patients undergoing bilateral cataract surgery were enrolled and randomly assigned to one of two groups: the bifocal group, including 30 eyes implanted with the bifocal diffractive IOL AT LISA 801 (Carl Zeiss Meditec, Jena, Germany), and the trifocal group, including eyes implanted with the trifocal diffractive IOL AT LISA tri 839 MP (Carl Zeiss Meditec). Analysis of visual and refractive outcomes, contrast sensitivity, ocular aberrations (OPD-Scan III; Nidek, Inc., Gagamori, Japan), and defocus curve were performed during a 3-month follow-up period. RESULTS: No statistically significant differences between groups were found in 3-month postoperative uncorrected and corrected distance visual acuity ( P ⩾ .21). However, uncorrected, corrected, and distance-corrected near and intermediate visual acuities were significantly better in the trifocal group ( P < .01). No significant differences between groups were found in postoperative spherical equivalent ( P = .22). In the binocular defocus curve, the visual acuity was significantly better for defocus of −0.50 to −1.50 diopters in the trifocal group ( P ⩽ .04) and −3.50 to −4.00 diopters in the bifocal group ( P ⩽ .03). No statistically significant differences were found between groups in most of the postoperative corneal, internal, and ocular aberrations ( P ⩾ .31), and in contrast sensitivity for most frequencies analyzed ( P ⩾ .15). CONCLUSIONS: Trifocal diffractive IOLs provide significantly better intermediate vision over bifocal IOLs, with equivalent postoperative levels of visual and ocular optical quality. [ J Refract Surg. 2014;30(10):666–672.]Keywords
This publication has 17 references indexed in Scilit:
- Design and qualification of a diffractive trifocal optical profile for intraocular lensesJournal of Cataract & Refractive Surgery, 2011
- Optical analysis, reading performance, and quality-of-life evaluation after implantation of a diffractive multifocal intraocular lensJournal of Cataract & Refractive Surgery, 2011
- Initial results of trifocal diffractive IOL implantationGraefe's Archive for Clinical and Experimental Ophthalmology, 2010
- Intermediate visual function with different multifocal intraocular lens modelsJournal of Cataract & Refractive Surgery, 2010
- Visual function after bilateral implantation of apodized diffractive aspheric multifocal intraocular lenses with a +3.0 D additionJournal of Cataract & Refractive Surgery, 2009
- Visual acuity comparison of 2 models of bifocal aspheric intraocular lensesJournal of Cataract & Refractive Surgery, 2009
- Prospective study of the Acri.LISA bifocal intraocular lensJournal of Cataract & Refractive Surgery, 2007
- Prospective visual evaluation of apodized diffractive intraocular lensesJournal of Cataract & Refractive Surgery, 2007
- Visual axial PSF of diffractive trifocal lensesOptics Express, 2005
- Prospective study of the new diffractive bifocal intraocular lensEye, 1989