High-resolution detection of Brownian motion for quantitative Optical Tweezers experiments
Authors:
Preprint
Abstract
We have developed a new in situ method to calibrate optical tweezers experiments and simultaneously measure the size of the trapped particle or the viscosity of the surrounding fluid. The positional fluctuations of the trapped particle are recorded with a high-bandwidth photodetector. Next, we compute the mean-square displacement, as well as the velocity autocorrelation function of the sphere and compare it to the theory of Brownian motion including hydrodynamic memory effects. A careful measurement and analysis of the time scales characterizing the dynamics of the harmonically bound sphere fluctuating in a viscous medium then directly yields all relevant parameters. Finally, we test the method for different optical trap strengths, with different bead sizes and in different fluids, and we find excellent agreement with the values provided by the manufacturers. The proposed approach overcomes the most commonly encountered limitations in precision when analyzing the power spectrum of position fluctuations in the region around the corner frequency. These low frequencies are usually prone to errors due to drift, limitations in the detection and trap linearity as well as short acquisition times resulting in poor statistics. Furthermore, the strategy can be generalized to Brownian motion in more complex environments, provided the adequate theories are available.Keywords
All Related Versions
- Version 1, 2012-08-06, ArXiv
- Published version: Physical Review E, 86 (2), 021912.
This publication has 29 references indexed in Scilit:
- Resonances arising from hydrodynamic memory in Brownian motionNature, 2011
- Direct observation of the full transition from ballistic to diffusive Brownian motion in a liquidNature Physics, 2011
- Surface Forces and Drag Coefficients of Microspheres near a Plane Surface Measured with Optical TweezersLangmuir, 2007
- Applications of laser tweezers in complex fluid rheologyCurrent Opinion in Colloid & Interface Science, 2005
- New calibration method for position detector for simultaneous measurements of force constants and local viscosity in optical tweezersOptics Communications, 2004
- Three-dimensional high-resolution particle tracking for optical tweezers by forward scattered lightMicroscopy Research and Technique, 1999
- Tracer microrheology in complex fluidsCurrent Opinion in Colloid & Interface Science, 1998
- Two-Dimensional Tracking of ncd Motility by Back Focal Plane InterferometryBiophysical Journal, 1998
- Direct observation of kinesin stepping by optical trapping interferometryNature, 1993
- Application of the Langevin equation to fluid suspensionsJournal of Fluid Mechanics, 1975