A new enhanced Howland voltage controlled current source circuit for EIT applications
- 1 February 2011
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 327-330
Abstract
A current source is one of the most important circuits in Electrical Impedance Tomography (EIT) hardware systems for measuring conductive multiphase flows. One of its important and challenging features is to deliver a relatively large and constant current for a wide range of resistive loads, high signal to noise ratio, and a large bandwidth. This paper presents a new voltage controlled current source for EIT applications. It is based on Howland circuit and uses two similar current sources circuits operating in parallel. Furthermore, the paper suggests a General Inverter Circuit (GIC) to remove the effect of eventual stray capacitances which usually appear across the load. Experimental results using Multisim software indicates that the new suggested circuit presents a more stable impedance output over a range of 100 to 3000 Ω. Furthermore, it was found that the current remains stable for range of 100 to 6000 Ω.Keywords
This publication has 5 references indexed in Scilit:
- Floating voltage-controlled current sources for electrical impedance tomography2007
- Current Source for Multifrequency Broadband Electrical Bioimpedance Spectroscopy Systems. A Novel Approach2006
- Current source design for electrical impedance tomographyPhysiological Measurement, 2003
- A one op-amp current source for electrical impedance tomography2003
- A comparison of modified Howland circuits as current generators with current mirror type circuitsPhysiological Measurement, 2000