Computational evaluation of the thrombogenic potential of a hollow-fiber oxygenator with integrated heat exchanger during extracorporeal circulation
- 6 October 2012
- journal article
- Published by Springer Nature in Biomechanics and Modeling in Mechanobiology
- Vol. 13 (2) , 349-361
Abstract
The onset of thromboembolic phenomena in blood oxygenators, even in the presence of adequate anticoagulant strategies, is a relevant concern during extracorporeal circulation (ECC). For this reason, the evaluation of the thrombogenic potential associated with extracorporeal membrane oxygenators should play a critical role into the preclinical design process of these devices. This study extends the use of computational fluid dynamics simulations to guide the hemodynamic design optimization of oxygenators and evaluate their thrombogenic potential during ECC. The computational analysis accounted for both macro- (i.e., vortex formation) and micro-scale (i.e., flow-induced platelet activation) phenomena affecting the performances of a hollow-fiber membrane oxygenator with integrated heat exchanger. A multiscale Lagrangian approach was adopted to infer the trajectory and loading history experienced by platelet-like particles in the entire device and in a repetitive subunit of the fiber bundles. The loading history was incorporated into a damage accumulation model in order to estimate the platelet activation state (PAS) associated with repeated passes of the blood within the device. Our results highlighted the presence of blood stagnation areas in the inlet section that significantly increased the platelet activation levels in particles remaining trapped in this region. The order of magnitude of PAS in the device was the same as the one calculated for the components of the ECC tubing system, chosen as a term of comparison for their extensive diffusion. Interpolating the mean PAS values with respect to the number of passes, we obtained a straightforward prediction of the thrombogenic potential as a function of the duration of ECC.Keywords
This publication has 44 references indexed in Scilit:
- Improving Oxygenator Performance Using Computational Simulation and Flow Field‐Based ParametersArtificial Organs, 2010
- Device Thrombogenicity Emulator (DTE) − Design optimization methodology for cardiovascular devices: A study in two bileaflet MHV designsJournal of Biomechanics, 2010
- High-Shear Stress Sensitizes Platelets to Subsequent Low-Shear ConditionsAnnals of Biomedical Engineering, 2010
- Towards Non-thrombogenic Performance of Blood Recirculating DevicesAnnals of Biomedical Engineering, 2010
- Bubble Tracking Through Computational Fluid Dynamics in Arterial Line Filters for Cardiopulmonary BypassASAIO Journal, 2009
- Platelet Activation Due to Hemodynamic Shear Stresses: Damage Accumulation Model and Comparison to In Vitro MeasurementsASAIO Journal, 2008
- Comparison of the Hemodynamic and Thrombogenic Performance of Two Bileaflet Mechanical Heart Valves Using a CFD/FSI ModelJournal of Biomechanical Engineering, 2007
- Turbulence Modeling in Three-Dimensional Stenosed Arterial BifurcationsJournal of Biomechanical Engineering, 2006
- Effect of biopassive and bioactive surface‐coatings on the hemocompatibility of membrane oxygenatorsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2006
- Shear-induced activation of plateletsJournal of Biomechanics, 1979