Ballistocardiogram artifact removal from EEG signals using adaptive filtering of EOG signals
Authors:
- Soo Y LeeSLSoo Y Lee
- Tae S ParkTPTae S Park
- Tae-S KimTKTae-S Kim
- Min H ChoMCMin H Cho
- Young B AhnYAYoung B Ahn
- 29 September 2006
- journal article
- Published by IOP Publishing in Physiological Measurement
- Vol. 27 (11) , 1227-1240
Abstract
We estimated ballistocardiogram (BCG) components in EEG signals recorded inside an MRI magnet using the electro-oculogram (EOG) signals recorded simultaneously with the EEG signals. Since the EOG signals are measured near the EEG measuring points, it is thought that the BCG components in the EOG signals resemble the BCG components in the EEG signals. To estimate the BCG components in the EEG signals, we applied the Kalman filter to the EOG and EEG signals recorded inside a 3.0 T MRI magnet. After removing the estimated BCG components from the EEG signals, we extracted the visual-evoked potentials (VEPs) from the BCG-removed EEG signals. To validate the efficacy of Kalman filtering in the BCG artifact removal, we have compared three types of VEPs of eight healthy subjects: one extracted from the raw EEG signals measured outside the magnet and the others extracted from the BCG-removed EEG signals measured inside the magnet. The BCG artifacts have been removed with Kalman filtering as well as with the conventional BCG template subtraction method for the sake of comparison. No significant difference in waveforms, latencies and amplitudes has been found between the two types of VEPs extracted from the two kinds of BCG-removed EEG signals.Keywords
This publication has 20 references indexed in Scilit:
- Visual evoked potentials recovered from fMRI scan periodsHuman Brain Mapping, 2005
- Visual evoked potentials may be recorded simultaneously with fMRI scanning: A validation studyHuman Brain Mapping, 2005
- Removal of ocular artifacts from electro-encephalogram by adaptive filteringMedical & Biological Engineering & Computing, 2004
- Motion and Ballistocardiogram Artifact Removal for Interleaved Recording of EEG and EPs during MRINeuroImage, 2002
- Acquiring simultaneous EEG and functional MRIClinical Neurophysiology, 2000
- A Method for Removing Imaging Artifact from Continuous EEG Recorded during Functional MRINeuroImage, 2000
- Integration of High Resolution EEG and Functional Magnetic Resonance in the Study of Human Movement-Related PotentialsMethods of Information in Medicine, 2000
- Spatiotemporal Activity of a Cortical Network for Processing Visual Motion Revealed by MEG and fMRIJournal of Neurophysiology, 1999
- Visual evoked potential (VEP) measured by simultaneous 64-channel EEG and 3T fMRINeuroReport, 1999
- Identification of EEG Events in the MR Scanner: The Problem of Pulse Artifact and a Method for Its SubtractionNeuroImage, 1998