Value of the unipolar electrogram in the diagnosis of right ventricular perforation following pacemaker implantation
Authors:
- J MacdonaldJMJ Macdonald
- D KellyDKD Kelly
- J WaktareJWJ Waktare
Open Access
Abstract
Ventricular depolarisation normally occurs from endocardium to epicardium. Endocardial lead positions therefore produce a predominantly negative unipolar electrogram (panel C) and epicardial sites result in a predominantly positive electrogram (panel A). The endocardial electrogram at the repositioned site (panel C) shows a small initial R wave, which is normal for an apical endocardial right ventricular lead position. This is because during normal intrinsic conduction the earliest site of ventricular activation is the septal aspect of the left ventricular endocardium and septal activation then occurs left to right.Keywords