Metal artifact reduction using virtual monochromatic images for patients with pedicle screws implants on CT
Authors:
- 13 June 2015
- journal article
- Published by Springer Nature in European Spine Journal
- Vol. 25 (6) , 1754-1763
Abstract
To evaluate the effect of spectral CT for metal artifact reduction in patients with pedicle screw. 45 patients with 119 pairs of pedicle screws underwent spectral CT examination. One set of conventional (140 kVp) polychromatic image and nine sets of virtual monochromatic images (60–140 keV) were obtained. The standard deviation (SD) of CT number in 12 locations around the implant and 1 on homogenous fat was measured to generate artifact index (AI). Objective assessment including AI, CT number and SD value was performed with independent t test and paired sample t test. Two radiologists independently reviewed the image quality, and Wilcoxon signed-rank test and kappa analysis were used for the subjective scores of image quality. The artifact index (AI) of all the regions decreased as keV increased. AIs of 100–140 keV were lower than that of 140 kVp images. At 120 keV there was no significant difference in CT numbers of psoas major muscle and vertebral canal between pedicle screw level and pedicle level, but a significant difference in SD value was determined between the two levels. The subjective scores at 100–140 keV were higher than the images at 140 kVp, and the highest subjective score of two observers and excellent interobserver agreement were found at 120 keV (κ = 0.889). Virtual monochromatic images at high-energy levels have a well-concordant effect of removing metal artifacts, and 120 keV monochromatic images provided an accurate CT number and good subjective score.Keywords
This publication has 19 references indexed in Scilit:
- Metal artifacts reduction using monochromatic images from spectral CT: Evaluation of pedicle screws in patients with scoliosisEuropean Journal of Radiology, 2013
- Spectral CT with Metal Artifacts Reduction Software for Improvement of Tumor Visibility in the Vicinity of Gold Fiducial MarkersRadiology, 2012
- Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction softwareEuropean Radiology, 2012
- Precision of the measurement of CT numbers: comparison of dual-energy CT spectral imaging with fast kVp switching and conventional CT with phantomsJapanese Journal of Radiology, 2011
- Monoenergetic Imaging of Dual-energy CT Reduces Artifacts from Implanted Metal Orthopedic Devices in Patients with FacturesAcademic Radiology, 2011
- New Techniques in Lumbar Spinal Instrumentation: What the Radiologist Needs to KnowRadiology, 2011
- Virtual Monochromatic Spectral Imaging with Fast Kilovoltage Switching: Improved Image Quality as Compared with That Obtained with Conventional 120-kVp CTRadiology, 2011
- Accuracy of pedicle screw placement in the lumbosacral spine using conventional technique: computed tomography postoperative assessment in 102 consecutive patientsJournal of Neurosurgery: Spine, 2010
- Artifacts in CT: Recognition and AvoidanceRadioGraphics, 2004
- Comparative Results Between Conventional and Computer-Assisted Pedicle Screw Installation in the Thoracic, Lumbar, and Sacral SpineSpine, 2000