Application of European protocol in the evaluation of contrast-to-noise ratio and mean glandular dose for two digital mammography systems
Authors:
- 18 February 2008
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 129 (1-3) , 231-236
Abstract
The performance of two digital mammography systems, Agfa CR75 and CRMM3 computed radiography (CR) and IMS Giotto MD direct digital radiography (DR), was assessed by applying a method recommended in the European protocol for quality control in mammography screening. The contrast-to-noise ratio (CNR) and mean glandular dose (MGD) values were measured and contrast detail (CD) analysis was performed. The CNRs for system CR were 21.9, 12.9, 9.5, 8.8, 7.4, 5.5 and 4.4 for 2, 3, 4, 4.5, 5, 6 and 7-cm polymenthylmethacrylate (PMMA) thickness, respectively. The respective CNRs for system DR were 10.4, 8.8, 6.3, 7.3, 7.2, 6.4 and 6.54. For the same phantom thickness sequence, the MGDs were 0.7, 1.1, 1.3, 1.6, 1.9, 2.5 and 3.4 mGy for system CR, whereas they were 0.7, 1.2, 1.1, 1.3, 1.8, 3.5 and 3.9 mGy for system DR. The CNR and MGD results satisfactorily correlate with CD analysis results. The MGD values compare well with the values recommended in the European protocol. Despite being simple, CNR and MGD can provide an effective system for performance assessment and constancy checks for related optimisations.Keywords
This publication has 11 references indexed in Scilit:
- Experimental investigation on the choice of the tungsten/rhodium anode/filter combination for an amorphous selenium-based digital mammography systemEuropean Radiology, 2007
- X‐ray spectrum optimization of full‐field digital mammography: Simulation and phantom studyMedical Physics, 2006
- An image quality comparison of standard and dual‐side read CR systems for paediatric radiologyMedical Physics, 2006
- Digital mammography: current state and future aspectsEuropean Radiology, 2005
- A framework for optimising the radiographic technique in digital X-ray imagingRadiation Protection Dosimetry, 2005
- Digital MammographyRadiology, 2005
- Experimental investigation of the dose and image quality characteristics of a digital mammography imaging systemMedical Physics, 2003
- Mammography Dose in Relation to Body Mass Index, Race, and Menopausal StatusRadiation Protection Dosimetry, 2002
- X-ray imaging with amorphous selenium: Optimal spectra for digital mammographyMedical Physics, 1996
- A model for optimization of spectral shape in digital mammographyMedical Physics, 1994