Magnetic Design of a Half-Size 5 T High-Temperature Superconducting Coil for MRI
- 11 April 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Applied Superconductivity
- Vol. 29 (5) , 1-5
Abstract
We have been developing a liquid-helium-free magnet using High-Temperature Superconducting (HTS) wire for MRI. The aim of this research and development is to commercialize a 3-T HTS magnet for whole body MRI which has the characteristics of being liquid-helium-free, small, and light-weight. HTS coil requires tolerance of a high magnetic field, 7 T, and it must have high current density, 200 A/mm 2 , for developing a small and light weight magnet used for 3-T MRI. A half-size 5-T HTS magnet will be manufactured in order to demonstrate the tolerance of high magnetic field and high current density of the coil, which are equivalent to a whole-body 3-T MRI magnet in a half-size HTS magnet. In this paper, we have investigated the arrangement of coils of a half-size 5-T coil. The designed coil consists of 124 double-pancake coils, whose inner diameter is 560 mm, outer diameter is 1200 mm, and axial length is 1100 mm. The uniformity of designed magnetic field is under 2.0 ppm (250 mmDSV).Keywords
Funding Information
- New Energy and Industrial Technology Development Organization
This publication has 9 references indexed in Scilit:
- Design and Manufacture of Half-Size 3-T High-Temperature Superconducting Magnet for MRIIEEE Transactions on Applied Superconductivity, 2018
- A compact 3 T all HTS cryogen-free MRI systemSuperconductor Science and Technology, 2017
- Towards liquid-helium-free, persistent-mode MgB 2 MRI magnets: FBML experienceSuperconductor Science and Technology, 2017
- Research and Development of the High Stable Magnetic Field ReBCO Coil System Fundamental Technology for MRIIEEE Transactions on Applied Superconductivity, 2016
- Conductors for commercial MRI magnets beyond NbTi: requirements and challengesSuperconductor Science and Technology, 2016
- Characteristics of REBCO Coated Conductors for 25 T Cryogen-Free Superconducting MagnetIEEE Transactions on Applied Superconductivity, 2014
- Newly Designed 3 T MRI Magnet Wound With Bi-2223 Tape ConductorsIEEE Transactions on Applied Superconductivity, 2013
- Advances in Whole-Body MRI MagnetsIEEE Transactions on Applied Superconductivity, 2010
- Artificial pinning center technology to enhance vortex pinning in YBCO coated conductorsSuperconductor Science and Technology, 2009