A comprehensive model for x-ray projection imaging system efficiency and image quality characterization in the presence of scattered radiation
Authors:
- P MonninPMP Monnin
- F R VerdunFVF R Verdun
- H BosmansHBH Bosmans
- S Rodríguez PérezSPS Rodríguez Pérez
- N W MarshallNMN W Marshall
Open Access
- 23 June 2017
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 62 (14) , 5691-5722
Abstract
This work proposes a method for assessing the detective quantum efficiency (DQE) of radiographic imaging systems that include both the x-ray detector and the antiscatter device. Cascaded linear analysis of the antiscatter device efficiency (DQEASD) with the x-ray detector DQE is used to develop a metric of system efficiency (DQEsys); the new metric is then related to the existing system efficiency parameters of effective DQE (eDQE) and generalized DQE (gDQE). The effect of scatter on signal transfer was modelled through its point spread function (PSF), leading to an x-ray beam transfer function (BTF) that multiplies with the classical presampling modulation transfer function (MTF) to give the system MTF. Expressions are then derived for the influence of scattered radiation on signal-difference to noise ratio (SDNR) and contrast-detail (c-d) detectability. The DQEsys metric was tested using two digital mammography systems, for eight x-ray beams (four with and four without scatter), matched in terms of effective energy. The model was validated through measurements of contrast, SDNR and MTF for poly(methyl)methacrylate thicknesses covering the range of scatter fractions expected in mammography. The metric also successfully predicted changes in c-d detectability for different scatter conditions. Scatter fractions for the four beams with scatter were established with the beam stop method using an extrapolation function derived from the scatter PSF, and validated through Monte Carlo (MC) simulations. Low-frequency drop of the MTF from scatter was compared to both theory and MC calculations. DQEsys successfully quantified the influence of the grid on SDNR and accurately gave the break-even object thickness at which system efficiency was improved by the grid. The DQEsys metric is proposed as an extension of current detector characterization methods to include a performance evaluation in the presence of scattered radiation, with an antiscatter device in place.Keywords
This publication has 55 references indexed in Scilit:
- A comparison of digital radiography systems in terms of effective detective quantum efficiencyMedical Physics, 2012
- Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology2012
- The effect of scatter and glare on image quality in contrast‐enhanced breast imaging using an full‐field flat panel detectorMedical Physics, 2009
- Scatter rejection in multislit digital mammographyMedical Physics, 2006
- Validation of MTF measurement for digital mammography quality controlMedical Physics, 2005
- Scatter/primary in mammography: Comprehensive resultsMedical Physics, 2000
- The effect of the antiscatter grid on full-field digital mammography phantom images.Journal of Digital Imaging, 1999
- An accurate method for computer‐generating tungsten anode x‐ray spectra from 30 to 140 kVMedical Physics, 1997
- Studies of performance of antiscatter grids in digital radiography: Effect on signal‐to‐noise ratioMedical Physics, 1990
- Characterization of the point spread function and modulation transfer function of scattered radiation using a digital imaging systemMedical Physics, 1986