Interatrial groove tear: An unusual complication of balloon atrial septostomy
Authors:
- 1 April 1983
- journal article
- case report
- Published by Springer Nature in Pediatric Cardiology
- Vol. 4 (2) , 149-150
Abstract
The authors present an unusual complication of the balloon atrial septostomy procedure performed in a neonate with D-transposition of the great arteries. Cardiac tamponade developed shortly after the balloon atrial septostomy procedure and the infant was found to have a tear in the superior aspect of the left atrium, parallel to the interatrial groove. The tear was successfully sutured and may have been caused by the greater pull-back force needed if the catheter balloon is maximally distended for the first pull-back.Keywords
This publication has 7 references indexed in Scilit:
- Percutaneous balloon atrial septostomy in infants with transposition of the great arteriesAmerican Heart Journal, 1976
- Thrombosis of the inferior vena cava following balloon septostomy in transposition of the great arteriesAmerican Heart Journal, 1971
- Balloon Embolization During Atrial SeptostomyCirculation, 1970
- The complications of balloon atrioseptostomyThe Journal of Pediatrics, 1970
- Inability to deflate catheter balloon: A complication of balloon atrial septostomyThe Journal of Pediatrics, 1970
- Creation of an atrial septal defect without thoracotomy. A palliative approach to complete transposition of the great arteries.1966
- Creation of an Atrial Septal Defect Without ThoracotomyJAMA, 1966