Insufficient mixing of thawed serum samples leading to erroneous results – experience from a field study and use of a correction procedure
- 18 December 2019
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 80 (2) , 99-105
Abstract
Incorrect analysis results that are close to expected might not be recognized in scientific studies or routine patient care. In two field studies we obtained unexpected results in a large number of samples. The present study aimed to identify the source of error in the samples from these studies and to validate a method to obtain correct results. Pre-analytical procedures were scrutinized, giving no indications of inappropriate pre-analytical sample handling in the field or during transport in a tropical climate. Using a new set of samples from volunteers in simulation experiments, we observed the known concentration gradient of analytes sampled in gel as well as plain tubes after freezer storage and thawing. Experiments demonstrated that mixing of samples by vortexing alone was not sufficient to disrupt the gradient formed by freezing and thawing, which appeared to cause the problem encountered when we in field studies analyzed and biobanked large sample sets by robot pipetting. A correction procedure was introduced, in which the obtained value of an analyte was multiplied by a correction factor calculated for each sample using the expected sodium level (140 mmol/L) divided by the measured sodium value. When it was validated on results from the simulation experiments, we repeatedly found that the correction lead to results very close to true values for analytes of different size and charge. Usefulness of the procedure was demonstrated when applied to a large set of field study results.Keywords
This publication has 14 references indexed in Scilit:
- Kidney function in sugarcane cutters in Nicaragua – A longitudinal study of workers at risk of Mesoamerican nephropathyEnvironmental Research, 2016
- Intervention to reduce heat stress and improve efficiency among sugarcane workers in El Salvador: Phase 1Occupational and Environmental Medicine, 2016
- Heat Stress Nephropathy From Exercise-Induced Uric Acid Crystalluria: A Perspective on Mesoamerican NephropathyAmerican Journal of Kidney Diseases, 2016
- Resolving the Enigma of the Mesoamerican Nephropathy: A Research Workshop SummaryAmerican Journal of Kidney Diseases, 2014
- An epidemic of chronic kidney disease in Central America: an overviewJournal of Epidemiology and Community Health, 2012
- Standard Preanalytical Coding for Biospecimens: Review and Implementation of the Sample PREanalytical Code (SPREC)Biopreservation and Biobanking, 2012
- Internal standards in the quantitative determination of protein biopharmaceuticals using liquid chromatography coupled to mass spectrometryJournal of Chromatography B, 2012
- Effect of Blood Collection Tubes on Total Triiodothyronine and Other Laboratory AssaysClinical Chemistry, 2005
- Current databases on biological variation: pros, cons and progressScandinavian Journal of Clinical and Laboratory Investigation, 1999
- Analytical error due to concentration gradients in frozen and thawed samplesClinica Chimica Acta, 1973