An improved intraoral transverse loop coil design for high‐resolution dental MRI
Open Access
- 23 June 2023
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 90 (4) , 1728-1737
Abstract
Purpose: To improve intraoral transverse loop coil design for high‐resolution dental MRI.Methods: The transverse intraoral loop coil (tLoop) was modified (mtLoop) by overlapping the feed port conductors, bending the posterior section, introducing a parallel plate capacitor, optimizing the insulation thickness, and using it in receive‐only mode. In addition, an MR‐silent insulation was introduced. The performances of the mtLoop and tLoop coils were compared in terms of sensitivity, image SNR, and eddy currents using electromagnetic simulations and MRI measurements at 3T.Results: The receive‐only mode of the mtLoop increases the sensitivity at the apices of the roots, and the overlapped feed port design eliminated signal voids along the incisors. The bent posterior section with the parallel plate capacitor reduced the unwanted signal of the tongue by a factor of 2.3 in the selected region off interest and lowered the eddy currents by 10%. The proposed new coil provided higher SNR by elevenfold and 2.5‐fold at the incisors and apices of the molar roots within the selected regions of interest, respectively, in the experiments, as well as improved comfort. Optimal insulation thickness was determined as 1 mm. With the mtLoop, a (250 μm)3 isotropic resolution of the dental arch could be realized using a UTE sequence within 2 min total acquisition time. A T2‐SPACE protocol with (350 μm)2 in‐plane resolution was also demonstrated.Conclusion: The proposed new coil offers higher SNR at the incisors and apices of the molar roots, less unwanted signals from tongue, lower eddy currents, and improved patient comfort.Keywords
This publication has 30 references indexed in Scilit:
- Dental MRI using wireless intraoral coilsScientific Reports, 2016
- Optimized 14 + 1 receive coil array and position system for 3D high-resolution MRI of dental and maxillomandibular structuresDentomaxillofacial Radiology, 2016
- Dental MRI using a dedicated RF-coil at 3 TeslaJournal of Cranio-Maxillofacial Surgery, 2015
- Intraoral approach for imaging teeth using the transverse B 1 field components of an occlusally oriented loop coilMagnetic Resonance in Medicine, 2013
- Ultrashort echo time (UTE) MRI for the assessment of caries lesionsDentomaxillofacial Radiology, 2013
- In vivo MRI‐based dental impression using an intraoral RF receiver coilConcepts in Magnetic Resonance Part B, 2008
- ImageJ for MicroscopyBioTechniques, 2007
- Stream function approach for determining optimal surface currentsJournal of Computational Physics, 2003
- Image averaging of flexible fibrous macromolecules: The clathrin triskelion has an elastic proximal segmentJournal of Structural Biology, 1991
- A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1–100 MHz: Dependence on tissue type, NMR frequency, temperature, species, excision, and ageMedical Physics, 1984