Performance of a factory‐calibrated, real‐time continuous glucose monitoring system during elective abdominal surgery
Open Access
- 13 June 2020
- journal article
- research article
- Published by Wiley in Diabetes, Obesity and Metabolism
- Vol. 22 (9) , 1678-1682
Abstract
We assessed the performance of the factory‐calibrated, sixth‐generation continuous glucose monitoring (CGM) system Dexcom G6® (DexCom Inc., San Diego, California) during elective abdominal surgery. Twenty adults with (pre)diabetes undergoing abdominal surgery (>2 hours; 15 men, age 69 ± 13 years, glycated haemoglobin 53 ± 14 mmol/mol) wore the sensor from 1 week prior to surgery until hospital discharge. From induction of anaesthesia until 2 hours post‐surgery, reference capillary glucose values were obtained every 20 minutes using the Accu‐Chek® Inform II meter (Roche Diabetes Care, Mannheim, Germany). The primary endpoint was the mean absolute relative difference (ARD) between sensor and reference method during this period. In total, 1207 CGM/reference pairs were obtained. In the peri‐operative period (523 pairs), mean ± SD and median (interquartile range [IQR]) ARD were 12.7% ± 8.7% and 9.9 (6.3;15.9)%, respectively, and 67.4% of sensor readings were within International Organization of Standardization 15197:2013 limits. CGM overestimated reference glucose by 1.1 ± 0.8 mmol/L (95% limits of agreement −0.5;2.7 mmol/L). Clarke error grid zones A or B contained 99.2% of pairs (A: 78.8%; B: 20.4%). The median (IQR) peri‐operative sensor availability was 98.6 (95.9;100.0)%. No clinically significant adverse events occurred. In conclusion, the Dexcom G6 device showed consistent and acceptable accuracy during elective abdominal surgery, opening new avenues for peri‐operative glucose management.Keywords
This publication has 13 references indexed in Scilit:
- Poor glycemic control is a strong predictor of postoperative morbidity and mortality in patients undergoing vascular surgeryJournal of Vascular Surgery, 2019
- Accuracy of a Factory-Calibrated, Real-Time Continuous Glucose Monitoring System During 10 Days of Use in Youth and Adults with DiabetesDiabetes Technology & Therapeutics, 2018
- Influence of Diabetes on Trends in Perioperative Cardiovascular EventsDiabetes Care, 2018
- Accuracy of 2 Different Continuous Glucose Monitoring Systems in Patients Undergoing Cardiac SurgeryJournal of Diabetes Science and Technology, 2016
- Continuous glucose monitoring system in the operating room and intensive care unit: any difference according to measurement sites?Journal of Clinical Monitoring and Computing, 2015
- Feasibility and Precision of Subcutaneous Continuous Glucose Monitoring in Patients Undergoing CABG SurgeryJournal of Cardiothoracic and Vascular Anesthesia, 2014
- Continuous Glucose Monitoring in Patients Undergoing Cardiac SurgeryDiabetes Technology & Therapeutics, 2012
- Randomized Study of Basal-Bolus Insulin Therapy in the Inpatient Management of Patients With Type 2 Diabetes Undergoing General Surgery (RABBIT 2 Surgery)Diabetes Care, 2011
- Intensive Intraoperative Insulin Therapy versus Conventional Glucose Management during Cardiac SurgeryAnnals of Internal Medicine, 2007
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987