Optimal beam quality selection based on contrast-to-noise ratio and mean glandular dose in digital mammography
Authors:
- 12 October 2010
- journal article
- research article
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 33 (4) , 329-334
Abstract
The performance of a digital mammography system (Siemens Mammomat Novation) using different target/filter combinations and tube voltage has been assessed. The objective of this study is to optimize beam quality selection based on contrast-to-noise ratio (CNR) and mean glandular dose (MGD). Three composition of breast were studied with composition of glandular/adipose of 30/70, 50/50, and 70/30. CNR was measured using 2, 4 and 6 cm-thick simulated breast phantoms with an aluminium sheet of 0.1 mm thickness placed on top of the phantom. Three target/filter combinations, namely molybdenum/molybdenum (Mo/Mo), molybdenum/rhodium (Mo/Rh) and tungsten/rhodium (W/Rh) with various tube voltage and mAs were tested. MGD was measured for each exposure. For 50/50 breast composition, Mo/Rh combination with tube voltage 26 kVp is optimal for 2 cm-thick breast. W/Rh combination with tube voltage 27 and 28 kVp are optimal for 4 and 6 cm-thick breast, respectively. For both 30/70 and 70/30 breast composition, W/Rh combination is optimal with tube voltage 25, 26 and 27 kVp, respectively. From our study it was shown that there are potential of dose reduction up to 11% for a set CNR of 3.0 by using beam quality other than that are determined by AEC selection. Under the constraint of lowest MGD, for a particular breast composition, calcification detection is optimized by using a softer X-ray beam for thin breast and harder X-ray beam for thick breast. These experimental results also indicate that for breast with high fibroglandular tissues (70/30), the use of higher beam quality does not always increase calcification detection due to additional structured noise caused by the fibroglandular tissues itself.Keywords
This publication has 8 references indexed in Scilit:
- Optimization of exposure parameters in full field digital mammographyMedical Physics, 2008
- Effect of dose reduction on the detection of mammographic lesions: A mathematical observer model analysisMedical Physics, 2007
- Optimal beam quality selection in digital mammographyBritish Journal of Radiology, 2006
- A study of mean glandular dose during diagnostic mammography in Malaysia and some of the factors affecting itBritish Journal of Radiology, 2003
- Experimental investigation of the dose and image quality characteristics of a digital mammography imaging systemMedical Physics, 2003
- Optimization of technique factors for a silicon diode array full‐field digital mammography system and comparison to screen‐film mammography with matched average glandular doseMedical Physics, 2003
- Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocolPhysics in Medicine & Biology, 2000
- Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study.British Journal of Radiology, 2000