CT Assessment of Intraorbital Cable Movement of Electronic Subretinal Prosthesis in Three Different Surgical Approaches
Open Access
- 1 July 2021
- journal article
- research article
- Published by Association for Research in Vision and Ophthalmology (ARVO) in Translational Vision Science & Technology
- Vol. 10 (8) , 16
Abstract
Purpose: Electronic retinal implants restore some visual perception in patients blind from retinitis pigmentosa. Eye movements cause mechanical stress in intraorbital power supply cables leading to cable breaks. By using computer tomography (CT) scans at the extreme positions of the four cardinal gaze directions, this study determined in vivo, which of three surgical routing techniques results in minimal bending radius variation and favors durability. Methods: Nine patients received the first-generation subretinal implant Alpha IMS (Retina Implant AG, Reutlingen, Germany) in one eye. Three techniques for intraorbital cable routing were used (straight cable route (A), parabulbar loop (B), and encircling band (C)), each in three patients. All patients underwent computer tomography of the orbital region. The bending radius of the intraorbital cable was measured with the DICOM viewer Osirix v4.1.2 (Pixmeo SARL, Bernex, Switzerland) and served as indicator for mechanical stress. Results: Average bending radius variation was 87% for method A, 11% for method B, and 16% for method C. Methods A and B (P = 0.001) and methods A and C (P = 0.001) differed significantly, while method B and C showed no statistical difference (P = 0.074). Conclusions: Compared to straight routes, arcuated cable routes significantly reduce cable movement and bending. Due to an easier surgical procedure, a parabulbar loop is the preferred method to minimize bending radius variation and prolong survival time of electronic subretinal implants. Translational Relevance: CT analysis of cable bending of implanted medical devices allows to determine which surgical routing technique favors durability in vivo.Keywords
This publication has 19 references indexed in Scilit:
- Oculomotor behavior of blind patients seeing with a subretinal visual implantVision Research, 2016
- Control and Functions of Fixational Eye MovementsAnnual Review of Vision Science, 2015
- Subretinal Visual Implant Alpha IMS – Clinical trial interim reportVision Research, 2015
- The unsteady eye: an information-processing stage, not a bugTrends in Neurosciences, 2015
- Restoring Natural Sensory Feedback in Real-Time Bidirectional Hand ProsthesesScience Translational Medicine, 2014
- Functional Outcome in Subretinal Electronic Implants Depends on Foveal EccentricityInvestigative Ophthalmology & Visual Science, 2013
- Ventriculoperitoneal shunt surgery and the incidence of shunt revision in adult patients with hemorrhage-related hydrocephalusClinical Neurology and Neurosurgery, 2012
- Subretinal electronic chips allow blind patients to read letters and combine them to wordsProceedings Of The Royal Society B-Biological Sciences, 2010
- Extraocular surgery for implantation of an active subretinal visual prosthesis with external connections: feasibility and outcome in seven patientsBritish Journal of Ophthalmology, 2008
- Preservation of the Inner Retina in Retinitis PigmentosaArchives of Ophthalmology (1950), 1997