Abstract
Measuring cerebral blood flow noninvasively and with high sensitivity is critical for clinical applications such as measuring the oxygen metabolic rate1,2 and monitoring intracranial pressure.3,4 Furthermore, although neuroscience applications such as functional activation mapping5,6 and noninvasive brain–computer interface7,8 have been pursued primarily using functional magnetic resonance imaging and near-infrared spectroscopy (fNIRS), such applications could in principle benefit from functional cerebral blood flow measurements.9 – 11 Diffuse correlation spectroscopy (DCS)12 is a promising noninvasive optical technique for monitoring cerebral blood flow13,14 and for measuring cortex functional activation during finger tapping9 and visual stimulation10,11 tasks. DCS measures deep-tissue dynamics by coupling coherent light into the subject and measuring the fluctuations in the speckle field created by the light diffusing out of the subject.12,15,16 Increasing the source–detector separation (ρ) of DCS optodes increases the proportion of detected photons that travel beneath the scalp and skull, deep into the brain cortex.17 – 19 However, the increase in sensitivity to deep tissue comes at the expense of a decrease in the signal-to-noise ratio (SNR) of the measured signal. Indeed, fewer diffuse photons reach the detector when the detector is farther away from the source due to absorption, scattering, and radial spread from the source. To increase the brain sensitivity of DCS, we developed multispeckle DCS (mDCS), a method that can extend ρ without sacrificing SNR by performing thousands of independent speckle measurements in parallel.