Wireless powered implantable bio-sensor tag system-on-chip for continuous glucose monitoring
Authors:
- 1 November 2011
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 21634025,p. 193-196
Abstract
An implantable passive radio frequency identification (RFID) sensor tag system-on-chip (SoC) for glucose monitoring is developed in this paper. A high frequency RFID tag, a bio-sensor interface, as well as the electrodes are included in this chip. The signal of blood sugar level from the bio-sensor is detected and converted into digital data, and then the data is encrypted and transmitted to an RFID reader on top of the skin. To measure the weak current signals, a current readout circuit, a current splitter, an ADC and a potentiostat are designed in the chip. Correlated double sample (CDS) technology is used in the readout circuit to reduce the 1/f noise and the op-amp's offset. The chip is taped out in SMIC 0.13 μm CMOS process. Simulation results show that the system can measure weak current signals in the range of 10 fA-100 pA.Keywords
This publication has 9 references indexed in Scilit:
- Wireless Implantable Electronic Platform for Chronic Fluorescent-Based BiosensorsIEEE Transactions on Biomedical Engineering, 2011
- Low power implementation of hummingbird cryptographic algorithm for RFID tag2010
- A redox-enzyme-based electrochemical biosensor with a CMOS integrated bipotentiostat2009
- A Wireless-Implantable Microsystem for Continuous Blood Glucose MonitoringIEEE Transactions on Biomedical Circuits and Systems, 2009
- Managing Subthreshold Leakage in Charge-Based Analog Circuits With Low-$V_rm TH$Transistors by Analog T- Switch (AT-Switch) and Super Cut-off CMOS (SCCMOS)IEEE Journal of Solid-State Circuits, 2006
- On the design and characterization of femtoampere current-mode circuitsIEEE Journal of Solid-State Circuits, 2003
- Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33)The Lancet, 1998
- Circuit techniques for reducing the effects of op-amp imperfections: autozeroing, correlated double sampling, and chopper stabilizationProceedings of the IEEE, 1996
- Intensive insulin therapy prevents the progression of diabetic microvascular complications in Japanese patients with non-insulin-dependent diabetes mellitus: a randomized prospective 6-year studyDiabetes Research and Clinical Practice, 1995