Development of a Deep Learning Model to Estimate Anemia from Palpebral Conjunctiva Taken with a Portable Slit-Lamp Microscope
Authors:
- Yo NakaharaYNYo Nakahara
- Eisuke ShimizuESEisuke Shimizu
- Takahiro MizukamiTMTakahiro Mizukami
- Hiroki NishimuraHNHiroki Nishimura
- Shintaro NakayamaSNShintaro Nakayama
- Tetsuo IshikawaTetsuo Ishikawa
- Masatoshi HirayamaMHMasatoshi Hirayama
- Risa HokamaRHRisa Hokama
- Kazuhiro SakuradaKSKazuhiro Sakurada
- Kazuno NegishiKNKazuno Negishi
Open Access
- 17 July 2026
- journal article
- research article
- Published by MDPI AG in Bioengineering
- Vol. 13 (7) , 824
Abstract
Background: Anemia is a common systemic condition associated with adverse maternal, perioperative, and cardiovascular outcomes. Although timely screening is clinically important, diagnosis still relies on invasive blood testing. Palpebral conjunctival pallor has traditionally been used as a noninvasive indicator of anemia, but its diagnostic accuracy remains limited. This study aimed to develop and validate a deep learning system to estimate hemoglobin (Hb) concentration and screen for anemia using palpebral conjunctiva images captured with a smartphone-compatible slit-lamp microscope. Methods: In this prospective observational study, 225 Japanese participants (20–92 years) underwent conjunctival imaging and blood testing. Palpebral conjunctiva videos were obtained using the Smart Eye Camera. Video frames were processed using automated anterior-segment segmentation and conjunctiva extraction. A ConvNeXt-based regression model was trained to predict Hb values. Anemia was defined using sex-specific Hb thresholds. Results: From 225 videos, 53,776 frames were extracted, yielding 9903 quality-filtered conjunctiva images (training: 8082; test: 1821). Video-level predicted Hb values moderately correlated with measured Hb (r = 0.42). For anemia screening, frame-level analysis achieved an AUC of 0.75, with accuracy of 0.76, sensitivity of 0.71, and specificity of 0.79. Video-level aggregation achieved 69% accuracy. Conclusions: Deep learning analysis of palpebral conjunctiva images acquired with a portable slit-lamp microscope demonstrated the feasibility of non-invasive hemoglobin estimation and anemia screening. Although the proposed approach achieved moderate performance, further improvements in model accuracy and prospective multi-center validation are required before clinical implementation.Keywords
Funding Information
- Japan Agency for Medical Research and Development
- Japan Science and Technology Agency
- Japan Society for the Promotion of Science (JSPS) Kakenhi
- New Energy and Industrial Technology Development Organization
This publication has 40 references indexed in Scilit:
- Artificial intelligence for anemia screening, diagnosis, and managementCurrent Research in Translational Medicine, 2025
- Highly reliable personalized noninvasive hemoglobin estimation by using Vision Transformers and dual fine-tuningComputers in Biology and Medicine, 2025
- Approach to anaemia in gastrointestinal disease: A position paper by the ESPGHAN Gastroenterology CommitteeJournal of Pediatric Gastroenterology and Nutrition, 2025
- Non-invasive hemoglobin estimation from conjunctival images using deep learningMedical Engineering & Physics, 2023
- Meta-analysis of the association between preoperative anaemia and mortality after surgeryBritish Journal of Surgery, 2015
- Combined reflectance spectroscopy and stochastic modeling approach for noninvasive hemoglobin determination via palpebral conjunctivaPhysiological Reports, 2014
- Worldwide prevalence of anaemia, WHO Vitamin and Mineral Nutrition Information System, 1993–2005Public Health Nutrition, 2009
- Anemia and Mortality in Heart Failure PatientsJACC, 2008
- Non-Invasive Determination of Hemoglobin by Digital Photography of Palpebral ConjunctivaThe Journal of Emergency Medicine, 2007
- The Relation of Conjunctival Pallor to the Presence of AnemiaJournal of General Internal Medicine, 1997