Abstract
Purpose: The purpose of this study is to report a case of Hemodialysis Reliable Outflow (HeRO) device malfunction in the subclavian position secondary to costoclavicular impingement. Methods and Results: The electronic medical record was reviewed for the patient in question and pertinent imaging was collected and photographs were taken with the patient's consent. The patient presented with signs of outflow obstruction after a period of HeRO venous outflow component function. The outflow component was found to be crushed between the first rib and the clavicle at the costoclavicular junction. Conclusions: This case suggests that costoclavicular impingement should be considered as a mode of outflow failure with the HeRO outflow component in the subclavian position. Other central venous access points should be considered first, and the central veins accessed through the lower extremities are possible alternatives if this failure mode was to arise. First rib resection would also be a viable strategy to relieve impingement at this site.Keywords
This publication has 11 references indexed in Scilit:
- Ultrasound-guided infraclavicular axillary vein puncture is effective to avoid pinch-off syndrome: a long-term follow-up studySurgery Today, 2012
- Abstract No. 97: Evaluation of a two-stage approach for insertion of the hemodialysis reliable outflow (HeRO) vascular access device in patients with central venous occlusionJournal of Vascular and Interventional Radiology, 2012
- Costoclavicular Venous Decompression in Patients With Threatened Arteriovenous Hemodialysis AccessAnnals of Vascular Surgery, 2011
- Surgical Management of Hemodialysis-Related Central Venous Occlusive Disease: A Treatment AlgorithmAnnals of Vascular Surgery, 2011
- Central Vein Stenosis: A Common Problem in Patients on HemodialysisASAIO Journal, 2005
- Central venous catheter pinch-off and fracture: recognition, prevention and managementBone Marrow Transplantation, 2003
- Subclavian Stents and Stent-Grafts: Cause for Concern?Journal of Endovascular Therapy, 1999
- Catheter Fracture and Embolization from Totally Implanted Venous Access PortsAngiology, 1998
- The mechanisms of positional dysfunction of subclavian venous catheters.Radiology, 1996
- Pinch-off syndrome: a complication of implantable subclavian venous access devices.Radiology, 1990