Implementation of a target volume design function for intrafractional range variation in a particle beam treatment planning system
Authors:
- T InaniwaTIT Inaniwa
- K MikiKMK Miki
- T ShiraiTST Shirai
- K NodaKNK Noda
- 1 November 2014
- journal article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 87 (1043) , 20140233
Abstract
Treatment planning for charged particle therapy in the thoracic and abdominal regions should take account of range uncertainty due to intrafractional motion. Here, we developed a design tool (4Dtool) for the target volume [field-specific target volume (FTV)], which accounts for this uncertainty using four-dimensional CT (4DCT). Target and normal tissue contours were input manually into a treatment planning system (TPS). These data were transferred to the 4Dtool via the picture archiving and communication system (PACS). Contours at the reference phase were propagated to other phases by deformable image registration. FTV was calculated using 4DCT on the 4Dtool. The TPS displays FTV contours using digital imaging and communications in medicine files imported from the PACS. These treatment parameters on the CT image at the reference phase were then used for dose calculation on the TPS. The tool was tested in single clinical case randomly selected from patients treated at our centre for lung cancer. In this clinical case, calculation of dose distribution with the 4Dtool resulted in the successful delivery of carbon-ion beam at the reference phase of 95% of the prescribed dose to the clinical target volume (CTV). Application to the other phases also provided sufficient dose to the CTV. The 4Dtool software allows the design of the target volume with consideration to intrafractional range variation and is now in routine clinical use at our institution. Our alternative technique represents a practical approach to four-dimensional treatment planning within the current state of charged particle therapy.Keywords
This publication has 25 references indexed in Scilit:
- Effects of a difference in respiratory cycle between treatment planning and irradiation for phase-controlled rescanning and carbon pencil beam scanningBritish Journal of Radiology, 2013
- Four-dimensional measurement of intrafractional respiratory motion of pancreatic tumors using a 256 multi-slice CT scannerRadiotherapy and Oncology, 2009
- Field‐size effect of physical doses in carbon‐ion scanning using range shifter platesMedical Physics, 2009
- Retrospective Analysis of Artifacts in Four-Dimensional CT Images of 50 Abdominal and Thoracic Radiotherapy PatientsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Quantification and Visualization of Charged Particle Range VariationsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- GPU-based streaming architectures for fast cone-beam CT image reconstruction and demons deformable registrationPhysics in Medicine & Biology, 2007
- Optimization for fast‐scanning irradiation in particle therapyMedical Physics, 2007
- 4D Proton treatment planning strategy for mobile lung tumorsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Design of 4D treatment planning target volumesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Treatment Planning for Heavy-ion Radiotherapy: Biological Optimization of Multiple Beam PortsJournal of Radiation Research, 2001