Controlled release of dry reagents in porous media for tunable temporal and spatial distribution upon rehydration
Authors:
- 24 August 2012
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 12 (21) , 4321-4327
Abstract
Novel methods are demonstrated that enable controlled spatial and temporal rehydration of dried reagents in a porous matrix. These methods can be used in paper-based microfluidic assays to define reagent concentrations over time at zones downstream for improved performance, and can reduce costs by simplifying the manufacturing process with the use of a single porous substrate. First, the creation of uniform reagent pulses from patterned arrays of dried reagent is demonstrated. Second, reagents are stored dry in separate regions of the porous matrix so that they can be combined upon rehydration for immediate use in the device. Third, reagents are reconstituted sequentially from dry storage depots with tunable delivery times. Fourth, the total time for dissolution is varied to achieve a range of reagent delivery times to a downstream region. Finally, the utility of these control methods is demonstrated in the context of real-time reagent rehydration and mixing on a porous device.Keywords
This publication has 21 references indexed in Scilit:
- Two-dimensional paper networks: programmable fluidic disconnects for multi-step processes in shaped paperLab on a Chip, 2011
- Enhanced Sensitivity of Lateral Flow Tests Using a Two-Dimensional Paper Network FormatAnalytical Chemistry, 2011
- Chemical signal amplification in two-dimensional paper networksSensors and Actuators B: Chemical, 2010
- Transport in two-dimensional paper networksMicrofluidics and Nanofluidics, 2010
- Controlled reagent transport in disposable 2D paper networksLab on a Chip, 2010
- Immunogold–silver staining-on-a-chip biosensor based on cross-flow chromatographyJournal of Chromatography B, 2010
- Diagnostics for the Developing World: Microfluidic Paper-Based Analytical Devices2009
- Imbibition in Porous Membranes of Complex Shape: Quasi-stationary Flow in Thin Rectangular SegmentsLangmuir, 2009
- Multiplex Lateral-Flow Test Strips Fabricated by Two-Dimensional ShapingACS Applied Materials & Interfaces, 2008
- Patterned Paper as a Platform for Inexpensive, Low‐Volume, Portable BioassaysAngewandte Chemie International Edition, 2007