Does the interface between individual 3D acetabular implants and host bone influence the functional outcomes in patients with severe bone loss after revision surgery?
Authors:
- Valery Yu. MurylevVMValery Yu. Murylev
- Grigory A. KukovenkoGKGrigory A. Kukovenko
- Olga EfimenkoOEOlga Efimenko
- Aleksei V. MuzychenkovAMAleksei V. Muzychenkov
- Pavel M. ElizarovPEPavel M. Elizarov
- Alexander I. RudnevARAlexander I. Rudnev
- Semyon S. AlekseevSASemyon S. Alekseev
- Dmitrii O. GolubkinDGDmitrii O. Golubkin
Open Access
- 24 October 2024
- journal article
- research article
- Published by EDP Sciences in SICOT-J
- Vol. 10, 42
Abstract
Introduction: There is a wide range of commercially produced revision implants for adequate reconstruction of acetabular large bone defects today, however, it is not always possible to achieve long-term survival of these implants. There is an increasing number of scientific publications concerning the use of custom-designed 3D components, which make it possible not only to achieve stable fixation and connect the pelvic bones but also to restore hip joint biomechanics. Objectives: To evaluate the positioning of 3D acetabular implants after revision hip arthroplasty and its impact on clinical and functional outcomes. Methods: we analyzed results in 48 patients with bone defect types IIIA and IIIB Paprosky types, after revision hip arthroplasty. A prospective study was conducted from 2017 to 2023. Revision arthroplasty due to aseptic loosening of the components was performed in 30 cases and as a second stage of periprosthetic infection treatment in 18 cases. Results: We did not achieve a statistically significant difference when using additional flanges and clinical and functional results. In 2 cases we faced aseptic loosening in patients using flanges. In no case were we able to install an implant with 100% adherence to porous structure compared to preoperatively planned adherence. According to the WOMAC and VAS scales, increasing the contact area of the components showed a slight statistical difference in the improvement of clinical and functional results and the reduction of pain. Conclusions: When acetabular 3D components adhered to the bone by more than 68%, we did not register a single complication in the postoperative period, and acetabular 3D components adhered to the bone by less than 68%, a total of 8 (16.6%) complications were registered.Keywords
Funding Information
- Moscow Center for Innovative Technologies in Healthcare (grant 2212-1/22-2 NIR dated April 29, 2022)
This publication has 33 references indexed in Scilit:
- Acetabular custom-made implants for severe acetabular bone defect in revision total hip arthroplasty: a systematic review of the literatureArchives of orthopaedic and trauma surgery, 2020
- Comparative analysis of current 3D printed acetabular titanium implants3D Printing in Medicine, 2019
- Management of Paprosky type three B acetabular defects by custom-made components: early resultsInternational Orthopaedics, 2018
- Customized implants for acetabular Paprosky III defects may be positioned with high accuracy in revision hip arthroplastyInternational Orthopaedics, 2018
- The 2018 Definition of Periprosthetic Hip and Knee Infection: An Evidence-Based and Validated CriteriaThe Journal of Arthroplasty, 2018
- Outcomes of Custom Flange Acetabular Components in Revision Total Hip Arthroplasty and Predictors of FailureThe Journal of Arthroplasty, 2016
- Comparative Epidemiology of Revision Arthroplasty: Failed THA Poses Greater Clinical and Economic Burdens Than Failed TKAClinical Orthopaedics and Related Research, 2015
- Custom 3D-Printed Acetabular Implants in hip Surgery–Innovative Breakthrough or Expensive Bespoke Upgrade?HIP International, 2015
- Acetabular defect classification and surgical reconstruction in revision arthroplastyThe Journal of Arthroplasty, 1994
- Revision arthroplasty using an anti-protrusio cage for massive acetabular bone deficiency.1992