Closed-Loop Delivery Systems Versus Manually Controlled Administration of Total IV Anesthesia: A Meta-analysis of Randomized Clinical Trials
Authors:
- Laura PasinLPLaura Pasin
- Pasquale NardelliPNPasquale Nardelli
- Margherita PintaudiMPMargherita Pintaudi
- Massimiliano GrecoMGMassimiliano Greco
- Massimo ZambonMZMassimo Zambon
- Luca CabriniLCLuca Cabrini
- Alberto ZangrilloAZAlberto Zangrillo
- 1 February 2017
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 124 (2) , 456-464
Abstract
Bispectral Index Scale (BIS)-guided closed-loop delivery of anesthetics has been extensively studied. We performed a meta-analysis of all the randomized clinical trials comparing efficacy and performance between BIS-guided closed-loop delivery and manually controlled administration of total IV anesthesia. Scopus, PubMed, EMBASE, and the Cochrane Central Register of clinical trials were searched for pertinent studies. Inclusion criteria were random allocation to treatment and closed-loop delivery systems versus manually controlled administration of total IV anesthesia in any surgical setting. Exclusion criteria were duplicate publications and nonadult studies. Twelve studies were included, randomly allocating 1284 patients. Use of closed-loop anesthetic delivery systems was associated with a significant reduction in the dose of propofol administered for induction of anesthesia (mean difference [MD] = 0.37 [0.17–0.57], P for effect P for heterogeneity = 0.001, I 2 = 74%) and a significant reduction in recovery time (MD = 1.62 [0.60–2.64], P for effect P for heterogeneity = 0.06, I 2 = 47%). The target depth of anesthesia was preserved more frequently with closed-loop anesthetic delivery than with manual control (MD = −15.17 [−23.11 to −7.24], P for effect P for heterogeneity I 2 = 83%). There were no differences in the time required to induce anesthesia and the total propofol dose. Closed-loop anesthetic delivery performed better than manual-control delivery. Both median absolute performance error and wobble index were significantly lower in closed-loop anesthetic delivery systems group (MD = 5.82 [3.17–8.46], P for effect P for heterogeneity I 2 = 90% and MD = 0.92 [0.13–1.72], P for effect = 0.003, P for heterogeneity = 0.07, I 2 = 45%). When compared with manual control, BIS-guided anesthetic delivery of total IV anesthesia reduces propofol requirements during induction, better maintains a target depth of anesthesia, and reduces recovery time.Keywords
This publication has 29 references indexed in Scilit:
- Association of Perioperative Risk Factors and Cumulative Duration of Low Bispectral Index with Intermediate-term Mortality after Cardiac Surgery in the B-Unaware TrialAnesthesiology, 2010
- Anesthetic Depth Is a Predictor of MortalityAnesthesiology, 2010
- Comparison of closed loop vs. manual administration of propofol using the Bispectral index in cardiac surgeryActa Anaesthesiologica Scandinavica, 2009
- Mortality Within 2 Years After Surgery in Relation to Low Intraoperative Bispectral Index Values and Preexisting Malignant DiseaseAnesthesia & Analgesia, 2009
- The Accuracy and Clinical Feasibility of a New Bayesian-Based Closed-Loop Control System for Propofol Administration Using the Bispectral Index as a Controlled VariableAnesthesia & Analgesia, 2008
- Feasibility of closed-loop titration of propofol guided by the Bispectral Index for general anaesthesia inductionEuropean Journal of Anaesthesiology, 2006
- Titration of Propofol for Anesthetic Induction and Maintenance Guided by the Bispectral Index: Closed-loop versus Manual ControlAnesthesiology, 2006
- Performance Evaluation of Two Published Closed-loop Control Systems Using Bispectral Index MonitoringAnesthesiology, 2004
- Closed-loop Control of Anesthesia Using Bispectral IndexAnesthesiology, 2002
- Closed-loop Feedback Control of Methohexital Anesthesia by Quantitative EEG Analysis in HumansAnesthesiology, 1987