A Case of HeartMate 3 Outflow Graft Twisting with Extraluminal Thrombosis: Is Computed Tomography Angiography Helpful?
Authors:
- Karim SalebKSKarim Saleb
- Marianna ZagurovskayaMZMarianna Zagurovskaya
- Michael SekelaMSMichael Sekela
- Maya GuglinMGMaya Guglin
Open Access
- 1 January 2020
- journal article
- Published by Texas Medical Center Library in The VAD Journal
- Vol. 6 (1) , e2020611
Abstract
Twists in the outflow graft of the HeartMateTM 3 device (Abbott) have recently been described as a sporadic, late complication. We present a case with a unique combination of external compression of the HeartMate 3 outflow graft by a surgical scar compounded by thrombus formation in the space between the band relief and the outflow graft with associated twist of the outflow graft and severe flow limitation. Computed tomography angiogram (CTA) of the chest was suggestive of outflow graft thrombosis. Our case sheds additional light on the limited specificity of gated CTA in distinguishing the outflow graft twisting from thrombotic obstruction and kinking.Keywords
This publication has 7 references indexed in Scilit:
- Cardiac CT to Diagnose Left Ventricular Assist Device Outflow Graft TwistRadiology, 2019
- A tale of the twist in the outflow graft: An analysis from the MOMENTUM 3 trialThe Journal of Heart and Lung Transplantation, 2018
- CT Imaging of Complications Associated with Continuous-Flow Left Ventricular Assist Devices (LVADs)Seminars in Ultrasound, CT and MRI, 2017
- Evaluation and treatment of pump thrombosis and hemolysis2014
- Use of Gated Cardiac Computed Tomography Angiography in the Assessment of Left Ventricular Assist Device DysfunctionASAIO Journal, 2011
- The Role of Echocardiography and Other Imaging Modalities in Patients With Left Ventricular Assist DevicesJACC: Cardiovascular Imaging, 2010
- Noninvasive assessment of left ventricular assist devices with cardiovascular computed tomography and impact on managementThe Journal of Heart and Lung Transplantation, 2010