Impact of the vaginal applicator and dummy pellets on the dosimetry parameters of Cs‐137 brachytherapy source
Open Access
- 19 May 2011
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 12 (3) , 183-193
Abstract
In this study, dose rate distribution around a spherical pellet source, from a low-dose-rate (LDR) Selectron remote afterloading system used in gynecological brachytherapy, has been determined using experimental and Monte Carlo simulation techniques. Monte Carlo simulations were performed using MCNP4C code, for a single pellet source in water medium and Plexiglas, and measurements were performed in Plexiglas phantom material using LiF TLD chips. Absolute dose rate distribution and the dosimetric parameters, such as dose rate constant, radial dose functions, and anisotropy functions, were obtained for a single pellet source. In order to investigate the effect of the applicator and surrounding pellets on dosimetric parameters of the source, the simulations were repeated for six different arrangements with a single active source and five non-active pellets inside central metallic tubing of a vaginal cylindrical applicator. In commercial treatment planning systems (TPS), the attenuation effects of the applicator and inactive spacers on total dose are neglected. The results indicate that this effect could lead to overestimation of the calculated F(r,θ), by up to 7% along the longitudinal axis of the applicator, especially beyond the applicator tip. According to the results obtained in this study, in a real situation in treatment of patients using cylindrical vaginal applicator and using several active pellets, there will be a large discrepancy between the result of superposition and Monte Carlo simulations. PACS number: 87.53.Jw
Keywords
This publication has 17 references indexed in Scilit:
- A seed specific dose kernel method for low-energy brachytherapy dosimetryJournal of Applied Clinical Medical Physics, 2003
- On the validity of the superposition principle in dose calculations for intracavitary implants with shielded vaginal colpostatsMedical Physics, 2001
- Dosimetric characteristics of a new 125I brachytherapy sourceMedical Physics, 2000
- Dosimetric characteristics of the InterSource103 palladium brachytherapy sourceMedical Physics, 2000
- The American Brachytherapy Society Survey of Brachytherapy Practice for Carcinoma of the Cervix in the United StatesGynecologic Oncology, 1999
- High dose-rate brachytherapy treatment delivery: Report of the AAPM Radiation Therapy Committee Task Group No. 59Medical Physics, 1998
- Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43Medical Physics, 1995
- Technically Accurate Intracavitary Insertions Improve Pelvic Control and Survival among Patients with Locally Advanced Carcinoma of the Uterine CervixGynecologic Oncology, 1994
- Shielding effects of Selectron applicator and pellets on isodose distributionsRadiotherapy and Oncology, 1991
- Dose calculations about shielded gynecological colpostatsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1990