Theoretical performance and clinical evaluation of transverse tripolar spinal cord stimulation
Authors:
- J.J. StruijkJSJ.J. Struijk
- J. HolsheimerJHJ. Holsheimer
- G.H.J. SpincemailleGSG.H.J. Spincemaille
- F.L.H. GielenFGF.L.H. Gielen
- R. HoekemaRHR. Hoekema
- 1 January 1998
- journal article
- case report
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Rehabilitation Engineering
- Vol. 6 (3) , 277-285
Abstract
A new type of spinal cord stimulation electrode, providing contact combinations with a transverse orientation, is presented. Electrodes were implanted in the cervical area (C4-C5) of two chronic pain patients and the stimulation results were subsequently simulated with a computer model consisting of a volume conductor model and active nerve fiber models. For various contact combinations a good match was obtained between the modeling results and the measurement data with respect to load resistance (less than 20% difference), perception thresholds (16% difference), asymmetry of paresthesia (significant correlation) and paresthesia distributions (weak correlation). The transversally oriented combinations provided the possibility to select either a preferential dorsal column stimulation, a preferential dorsal root stimulation or a mixed stimulation. The (a)symmetry of paresthesia could largely be affected in a predictable way by the selection of contact combinations as well. The transverse tripolar combination was shown to give a higher selectivity of paresthesia than monopolar and longitudinal dipolar combinations, at the cost of an increased current (more than twice).Keywords
This publication has 22 references indexed in Scilit:
- Time Dependent Variations of Stimulus Requirements in Spinal Cord Stimulation for Angina PectorisPacing and Clinical Electrophysiology, 1997
- Excitation of dorsal root fibers in spinal cord stimulation: a theoretical studyIEEE Transactions on Biomedical Engineering, 1993
- Paresthesia thresholds in spinal cord stimulation: a comparison of theoretical results with clinical dataIEEE Transactions on Rehabilitation Engineering, 1993
- Mapping of sensory responses to epidural stimulation of the intraspinal neural structures in manJournal of Neurosurgery, 1993
- Recruitment of dorsal column fibers in spinal cord stimulation: influence of collateral branchingIEEE Transactions on Biomedical Engineering, 1992
- How Do Geometrie Factors Influence Epidural Spinal Cord Stimulation?Stereotactic and Functional Neurosurgery, 1991
- Contact Combinations in Epidural Spinal Cord StimulationStereotactic and Functional Neurosurgery, 1991
- Central somatosensory conduction in man: Neural generators and interpeak latencies of the far-field components recorded from neck and right or left scalp and earlobesElectroencephalography and Clinical Neurophysiology, 1980
- A quantitative description of membrane currents in rabbit myelinated nerve.The Journal of Physiology, 1979
- Analysis of a Model for Excitation of Myelinated NerveIEEE Transactions on Biomedical Engineering, 1976