Comparison of invasive and non-invasive blood pressure monitoring during clinical anaesthesia in dogs
- 23 March 2010
- journal article
- research article
- Published by Springer Nature in Veterinary Research Communications
- Vol. 34 (3) , 217-227
Abstract
Monitoring blood pressure during anaesthesia is widely recommended in man and animals. The accuracy of any device used to measure blood pressure is an important consideration when selecting monitoring equipment, the ANSI/AAMI SP10 standard is widely cited in this respect in recent veterinary publications. Blood pressure was monitored using invasive and non-invasive techniques during clinical anaesthesia in 19 dogs. The results were compared using Bland-Altman analysis. The bias (and limits of agreement) between invasive and non-invasive measurement was 7.1 mmHg (+/−34.7) for systolic blood pressure, −1.8 mmHg (+/−27.4) for mean blood pressure and 6.9 mmHg (+/−27.5) for diastolic blood pressure. In a clinical setting the bias between invasive and non-invasive measurement techniques was similar or smaller than laboratory reports, however the limits of agreement were considerably wider suggesting that care should be exercised when interpreting NIBP values.Keywords
This publication has 24 references indexed in Scilit:
- Results of the Confidential Enquiry into Perioperative Small Animal Fatalities regarding risk factors for anesthetic-related death in dogsJournal of the American Veterinary Medical Association, 2008
- Clinical evaluation of the Surgivet V60046, a non invasive blood pressure monitor in anaesthetized dogsVeterinary Anaesthesia and Analgesia, 2008
- Blood pressure measurementContinuing Education in Anaesthesia Critical Care & Pain, 2007
- Accuracy of an oscillometric blood pressure monitor during phenylephrine-induced hypertension in dogsAmerican Journal of Veterinary Research, 2006
- Accuracy of Indirect Measurement of Blood Pressure in Neonatal FoalsJournal of Veterinary Internal Medicine, 2005
- Comparison of direct and indirect (oscillometric) measurements of arterial blood pressure in conscious dogsResearch in Veterinary Science, 1996
- Epidemiological study of blood Sressure in omestic dogsJournal of Small Animal Practice, 1996
- Doppler Ultrasonographic, Oscillometric Sphygmomanometric, and Photoplethysmographic Techniques for Noninvasive Blood Pressure Measurement in Anesthetized CatsJournal of Veterinary Internal Medicine, 1995
- The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systemsJournal Of Hypertension, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986