Wearable cuffless blood pressure monitoring devices: a systematic review and meta-analysis

Authors:

Top Cited Papers
Open Access
Abstract
High blood pressure (BP) is the commonest modifiable cardiovascular risk factor, yet its monitoring remains problematic. Wearable cuffless BP devices offer potential solutions; however, little is known about their validity and utility. We aimed to systematically review the validity, features and clinical use of wearable cuffless BP devices. We searched MEDLINE, Embase, IEEE Xplore and the Cochrane Database till December 2019 for studies that reported validating cuffless BP devices. We extracted information about study characteristics, device features, validation processes, and clinical applications. Devices were classified according to their functions and features. We defined devices with a mean systolic BP (SBP) and diastolic BP (DBP) biases of 2 95.4%) and DBP (pooled mean=1.16 mmHg, 95%CI -1.26, 3.58, I2 87.1%). Several cuffless BP devices are currently available using different technologies, offering the potential for continuous BP monitoring. The variation in standards and validation protocols limited the comparability of findings across studies and the identification of the most accurate device. Challenges such as validation using standard protocols and in real-life settings must be overcome before they can be recommended for uptake into clinical practice.