Abstract
Introduction: Non‐parallelism in factor assays can lead to incorrect factor activities. Parallelism can be assessed by calculating the coefficient of variation (CV) of results obtained on 3 dilutions of the same sample. Some authors have proposed that if there is 25 IU/dl (plus no interfering substances) all CVs were < 15%. One sample (low factor activities 10–15 IU/dl), had CVs > 15% in FII, FVII and FXII assays only. SR outside of 0.9–1.1 were seen in FII and FXII assays at different levels of clotting factor including some within the normal range. Non‐parallelism was detected more frequently with SR than CV for those with interfering substances.Conclusions: SR outside of 0.9–1.1 were seen in different levels of clotting factors, including samples which did not contain interfering substances. The target of 15% CV was a better discriminator than a SR for acceptance. When factor levels were reduced to around 10–15 IU/dl, a target 20 %CV was more appropriate than 15%. It might be appropriate for laboratories to assess locally whether their acceptance criteria need to be wider at low levels of clotting factors.