Inverse‐geometry volumetric CT system with multiple detector arrays for wide field‐of‐view imaging
Authors:
- Samuel R. MazinSMSamuel R. Mazin
- Josh Star‐LackJSJosh Star‐Lack
- N. Robert BennettNBN. Robert Bennett
- Norbert J. PelcNPNorbert J. Pelc
- 21 May 2007
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 34 (6Part1) , 2133-2142
Abstract
Current volumetric computed tomography (CT) methods require seconds to acquire a thick volume with high resolution. Inverse‐geometry CT (IGCT) is a new system geometry under investigation that is anticipated to be able to image a thick volume in a single gantry rotation with isotropic resolution and no cone‐beam artifacts. IGCT employs a large array of source spots opposite a smaller detector array. The in‐plane field of view (FOV) is primarily determined by the size of the source array, in much the same way that the FOV is determined by the size of the detector array in a conventional CT system. Thus, the size of the source array can be a limitation on the achievable FOV. We propose adding additional detector arrays, spaced apart laterally, to increase the in‐plane FOV while still using a modestly sized source array. We determine optimal detector placement to maximize the FOV while obtaining relatively uniform sampling. We also demonstrate low wasted radiation of the proposed system through design and simulation of a pre‐patient collimator. Reconstructions from simulated projection data show no artifacts when combining the data from the detector arrays. Finally, to demonstrate feasibility of the concept, an anthropomorphic thorax phantom containing a porcine heart was scanned on a prototype table‐top system. The reconstructed axial images demonstrate a in‐plane FOV using a source array.Keywords
This publication has 14 references indexed in Scilit:
- A prototype table-top inverse-geometry volumetric CT systemMedical Physics, 2006
- An inverse‐geometry volumetric CT system with a large‐area scanned source: A feasibility studyMedical Physics, 2004
- Multi‐slice helical CT: Scan and reconstructionMedical Physics, 1999
- Dual‐slice spiral versus single‐slice spiral scanning: Comparison of the physical performance of two computed tomography scannersMedical Physics, 1996
- Selection of a convolution function for Fourier inversion using gridding (computerised tomography application)IEEE Transactions on Medical Imaging, 1991
- Computed tomography scanning with simultaneous patient translationMedical Physics, 1990
- Spiral volumetric CT with single-breath-hold technique, continuous transport, and continuous scanner rotation.Radiology, 1990
- Cone-beam tomography: recent advances and a tutorial reviewOptical Engineering, 1990
- A Fast Sinc Function Gridding Algorithm for Fourier Inversion in Computer TomographyIEEE Transactions on Medical Imaging, 1985
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984