EVALUATION OF A NEW AUTOMATIC DEVICE FOR TAKING AND RECORDING BLOOD PRESSURES

Authors:
  • SHS W Hoobler
    S W Hoobler
  • BOB Oesterle
    B Oesterle
  • HEH Early
    H Early

  • 1 January 1976
    • journal article
    • research article
    • Vol. 88  (5) , 826-833
    • PMID:978045
Abstract
A new instrument capable of recording systolic and diastolic blood pressure automatically and repetitively is described, and its accuracy is equal to that obtained by a carefully trained observer among 53 patients with different levels of blood pressure. The instrument (BosoMat) is also more accurate than 2 others available, namely, the Arteriosonde Model 1213 and the SphygmoStat Model B-300. The standard deviation among 3 consecutive blood pressures on the same patient as determined by the new instrument or by the nurse using traditional methods is approximately 4 mm Hg.