Clinical and technical determinants of long-term performance of coronary sinus leads
Authors:
- Emanuele BertagliaEBEmanuele Bertaglia
- Francesco SolimeneFSFrancesco Solimene
- Michele AccogliMAMichele Accogli
- Assunta IulianoAIAssunta Iuliano
- Natale MarrazzoNMNatale Marrazzo
- Gabriele De Luca De MasiGMGabriele De Luca De Masi
- Alberto ScacciaASAlberto Scaccia
- Giovanni DonniciGDGiovanni Donnici
- Paola ChiarielloPCPaola Chiariello
- Pietro TurcoPTPietro Turco
- Antonio De SimoneASAntonio De Simone
- 1 January 2010
- journal article
- Published by Wolters Kluwer Health in Journal of Cardiovascular Medicine
- Vol. 11 (1) , 40-44
Abstract
Little is known about coronary sinus lead performance in patients with cardiac resynchronization therapy devices. We evaluated the impact of clinical and technical parameters on coronary sinus lead performance over long-term follow-up. From February 1999 to July 2004, 235 patients (181 men; mean age, 68 +/- 9 years; mean left ventricular ejection fraction, 26.5 +/- 6.5%; idiopathic dilated cardiomyopathy, 49%; ischemic, 48%; and other cause, 3%) underwent cardiac resynchronization therapy (pacemaker or defibrillator) implantation. On implantation, the only statistically significant difference was observed in mean pacing impedance, which was lower for unipolar leads than for bipolar leads (763 +/- 250 vs. 847 +/- 270 Omega, P = 0.02), and lower in patients with ischemic cardiomyopathy than in those with idiopathic cardiomyopathy (758 +/- 204 vs. 837 +/- 291 Omega, P = 0.03). After a mean follow-up of 41.7 +/- 14.7 months, a significant decrease was observed in mean pacing impedance (from 811 +/- 261 to 717 +/- 284 Omega, P = 0.0026) and mean R-wave amplitude (from 13.2 +/- 6.7 to 10.5 +/- 5.5 mV, P = 0.002), whereas the mean pacing energy threshold increased from 3.6 +/- 10.4 to 9.5 +/- 28 microJ (P = 0.004). On analysis of variance, unipolar lead (P = 0.016) and posterior coronary sinus position (P = 0.049) were related to a greater energy threshold increase. On multivariate analysis (Cox regression), only posterior coronary sinus lead position (P = 0.013) proved to be an independent predictor of long-term significant increase in the stimulation energy threshold. Over the long-term follow-up of coronary sinus leads, pacing impedance and R-wave amplitude decreased, whereas the energy threshold increased; unipolar leads and posterior lead location in the coronary sinus were related to a greater energy threshold increase.Keywords
This publication has 14 references indexed in Scilit:
- Implantation of cardiac resynchronization therapy systems in the CARE-HF trial: procedural success rate and safetyEP Europace, 2007
- Single-centre experience with coronary sinus lead stability and long-term pacing parametersEP Europace, 2007
- Long‐Term Performance of Coronary Sinus Leads Used for Cardiac Resynchronization TherapyPacing and Clinical Electrophysiology, 2007
- Safety of Transvenous Cardiac Resynchronization System Implantation in Patients With Chronic Heart FailureJACC, 2005
- Left Ventricular Lead Performance in Cardiac Resynchronization Therapy: Impact of Lead Localization and ComplicationsPacing and Clinical Electrophysiology, 2005
- Evolution of left and right ventricular capture thresholds over time in patients receiving cardiac re-synchronization therapyHeart Rhythm, 2005
- Effect of Left Ventricular Function on Long-Term Left Ventricular Pacing and Sensing ThresholdJournal of Interventional Cardiac Electrophysiology, 2003
- Six year experience of transvenous left ventricular lead implantation for permanent biventricular pacing in patients with advanced heart failure: technical aspectsHeart, 2001
- Performance of a New Steroid‐Eluting Coronary Sinus Lead Designed for Left Ventricular PacingPacing and Clinical Electrophysiology, 2000
- Transvenous biventricular pacing for heart failure: can the obstacles be overcome?The American Journal of Cardiology, 1999