External Magnet Displacement in Cochlear Implants

Authors:
  • ASAmanda Scott
    Amanda Scott
  • CPColleen Polite
    Colleen Polite
  • LLLawrence R. Lustig
    Lawrence R. Lustig

Abstract
To evaluate the complication of external magnet displacement in cochlear implant patients. Retrospective case review of patients at a tertiary academic medical center. Eleven patients were identified with postoperative magnetic retention difficulties after cochlear implantation. Skin flap thickness test, hair shaving over the magnet site, elastic headband usage, and skin flap thinning or revision were the primary interventions used. Qualitative reports of magnet retention difficulty noted in patient records during audiologic and device programming follow-up visits. Follow-up ranged from 2 to 4 years from the date of implantation. Interventions included initial conservative measures (e.g., shaving hair, wearing headband). Four patients required flap thinning surgery; of these, 3 showed marked improvement. Almost all patients in this series were overweight or obese. Patients with external magnet retention difficulties can present a challenge to users of cochlear implants and their providers. Conservative measures will alleviate the issue in many cases. Skin flap thinning is a viable option for those patients whose magnet retention difficulties do not resolve with conservative treatments. However, extra vigilance must be given to the skin flap in all cases to monitor the effects of both conservative and nonconservative measures because overcorrection may risk skin breakdown.