Gamma-camera uniformity as a function of energy and count-rate
Authors:
- A. HasmanAHA. Hasman
- 1 August 1976
- journal article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 49 (584) , 718-722
Abstract
Recently there has been some discussion on the dependence of the uniformity of a gamma-camera on the energy of the incident photons and on their count-rate (Hannan and Bessent, 1974; Jansson and Parker, 1975). Since in these cases only the results of a few investigations were reported it was considered worthwhile to examine this problem more thoroughly. This paper reports the results of experiments performed with a Picker Dynacamera 2C system. The Picker Dynacamera is coupled to a PDP-8E computer through the image enhancement system (interface to the computer developed by Picker). The PDP-8E computer has a memory of 16K making possible pictures with 100×100 elements. An extended arithmetic element performs fast multiply/divide manipulations which are important for fast corrections of non-uniformity. The data are stored on a 1·6 million word disc (RK/8E from Digital Equipment).Keywords
This publication has 2 references indexed in Scilit:
- Pitfalls in gamma camera field uniformity correctionBritish Journal of Radiology, 1975
- Pitfalls in gamma camera field uniformity correctionBritish Journal of Radiology, 1974