A method for calibrating coil constants by using the free induction decay of noble gases

Authors:

Abstract
We propose a precise method to calibrate the coil constants of spin-precession gyroscopes and optical atomic magnetometers. This method is based on measuring the initial amplitude of Free Induction Decay (FID) of noble gases, from which the π/2 pulse duration can be calculated, since it is inversely proportional to the amplitude of the π/2 pulse. Therefore, the coil constants can be calibrated by measuring the π/2 pulse duration. Compared with the method based on the Larmor precession frequency of atoms, our method can avoid the effect of the pump and probe powers. We experimentally validated the method in a Nuclear Magnetic Resonance Gyroscope (NMRG), and the experimental results show that the coil constants are 436.63±0.04 nT/mA and 428.94±0.02 nT/mA in the x and y directions, respectively.
Funding Information
  • National Natural Science Foundation of China (61227902, 61673041)
  • Ministry of Science and Technology of the People's Republic of China (2016YFB0501601)

This publication has 22 references indexed in Scilit: