Screw Fixation Compared to H-Locking Plate Fixation for First Metatarsocuneiform Arthrodesis: A Biomechanical Study
Authors:
- 1 November 2005
- journal article
- research article
- Published by SAGE Publications in Foot & Ankle International
- Vol. 26 (11) , 984-989
Abstract
Background: Several different techniques have been used for fixation of first metatarsocuneiform (MTC) joint arthrodesis, a standard treatment for arthritis, instability, and deformity of the MTC joint. Improved plating systems using locking designs are now available, but no studies have yet compared this construct with other methods. We compared load to failure with a locking plate design versus standard crossed-screw fixation. Methods: Ten matched pairs of fresh frozen cadaver feet were used. The bone density of each pair was measured with DEXA scanning. One foot of each pair was randomly assigned to have a dorsomedial Normed H titanium locking plate (Normed Medizin-Technik Vertriebs-GmbH, D-78501 Tuttlingen, Germany) applied to the first MTC joint. On the other foot of the pair, fixation of the first MTC joint was done with crossed ACE DePuy 4.0 (DePuy/Ace, Warsaw, IN) titanium cannulated screws. The first metatarsal and first cuneiform were then isolated and planted in an epoxy resin. The specimens were loaded to failure in a four-point bending configuration using a MTS Mini Bionix test frame (MTS Systems Corp., Eden Prairie, MN). Failure was defined as displacement of more than 3 mm at the arthrodesis site. The Student t-test was used to determine any observed differences, with significance set at p ≤ 0.05. Results: The mean maximal load to failure was 140.08 N (SD ± 77.1) for screw fixation alone and 58.09 N (SD ± 11.86) for the H-locking plate. This difference was statistically significant ( p = 0.008). The mean stiffness of the construct for screw fixation alone was 83.10 N/mm (SD ± 49.8) and 19.96 N/mm for the H-locking plate. This difference also was statistically significant ( p = 0. 0 04). Conclusion: Screw fixation for first MTC arthrodesis created a stronger and stiffer construct than did the H-locking plate. This was likely due to the mechanical design of the implants. Compression across the MTC joint could be applied with the screws, but the plate relied on a fixed angle design with no compression.Keywords
This publication has 9 references indexed in Scilit:
- Ligamentous Lisfranc Joint Injuries: A Biomechanical Comparison of Dorsal Plate and Transarticular Screw FixationFoot & Ankle International, 2005
- Deformity and Degeneration in the Hindfoot and Midfoot Joints of the Adult Acquired FlatfootFoot & Ankle International, 2003
- Modified Lapidus Procedure for the Treatment of Hypermobile Hallux ValgusFoot & Ankle International, 2000
- Mechanical Comparison of Biplanar Proximal Closing Wedge Osteotomy with Plantar Plate Fixation Versus Crescentic Osteotomy with Screw Fixation for the Correction of Metatarsus Primus VarusFoot & Ankle International, 1998
- Clinical, Quantitative Assessment of First Tarsometatarsal Mobility in the Sagittal Plane and Its Relation to Hallux Valgus DeformityFoot & Ankle International, 1994
- Deformity Correction and Arthrodesis of the Midfoot with a Medial Plate*Foot & Ankle, 1993
- Metatarsocuneiform Arthrodesis for Management of Hallux Valgus and Metatarsus Primus VarusFoot & Ankle, 1992
- Modified Lapidus Procedure for Hallux ValgusFoot & Ankle, 1989
- Hypermobility in Hallux ValgusFoot & Ankle, 1988