Preference judgments of artificial processed and hearing-aid transduced speech
Authors:
- 1 September 1999
- journal article
- research article
- Published by Acoustical Society of America (ASA) in The Journal of the Acoustical Society of America
- Vol. 106 (3) , 1566-1578
Abstract
In order to assess the relative importance of various signal processing algorithms and distortions on hearing-aid preference, male and female speech was manipulated in a number of ways and subsequently presented to normal-hearing and hearing-impaired subjects (the latter having a mild sensorineural high-frequency hearing loss). Signal manipulations were artificial (e.g., filtering, compression, peak clipping, or adding noise) or were actual dummy-head recordings of five different hearing aids. Listeners judged the sounds in a pairwise-comparison format. Their task was to indicate the “hearing aid” they would prefer assuming they had to wear it all day. The data were analyzed with multidimensional scaling techniques; Principal Components Analysis revealed that the first two dimensions on which preference judgments were based, can be interpreted as (1) intelligibility or clarity, and (2) distinction between signal distortion and added background distortion. Furthermore, the results showed that normal-hearing subjects generally preferred the original signal, whereas hearing-impaired subjects were inclined to choose the signals with a high-frequency emphasis. Severe band-pass filtering or low-frequency emphasis were disliked, as was to be expected. Surprisingly, however, a soft background noise (S/N ratio of 25 dB) was often among the least preferred of all signals. The differences in preference between the five hearing aids were small, but consistent. For hearing-impaired subjects, hearing-aid ordering could be accounted for by the amount of low-frequency cutoff; for normal-hearing subjects both high- and low-frequency cutoff played a role. Results of a retest experiment with normal-hearing subjects, about one year later, showed that subjects’ criteria remain remarkably stable.Keywords
This publication has 34 references indexed in Scilit:
- Effect of release time in compression hearing aids: Paired-comparison judgments of qualityThe Journal of the Acoustical Society of America, 1995
- Comparison of Two Hearing Aid Receiver-Amplifier Combinations Using Sound Quality JudgmentsEar & Hearing, 1995
- Optimal estimation of hearing-aid compression parametersThe Journal of the Acoustical Society of America, 1993
- The National Acoustic Laboratoriesʼ (NAL) New Procedure for Selecting the Gain and Frequency Response of a Hearing AidEar & Hearing, 1986
- Relative Effects of Low-Frequency Amplification on Syllable Recognition and Speech QualityEar & Hearing, 1986
- Effect of Low-Frequency Hearing Aid Response on Four Measures of Speech PerceptionEar & Hearing, 1984
- Note on the construction of digram-balanced Latin squares.Psychological Bulletin, 1969
- Multidimensional Analyses of Circuit Quality JudgmentsThe Journal of the Acoustical Society of America, 1969
- Measurement of Perceived Acoustic Quality of Sound-Reproducing Systems by Means of Factor AnalysisThe Journal of the Acoustical Society of America, 1966
- A law of comparative judgment.Psychological Review, 1927