A Wide-Field Micro-Computed Tomography Detector: Micron Resolution at Half-centimeter Scale
Authors:
- Maksim A. YakovlevMYMaksim A. Yakovlev
- Daniel J. VanselowDVDaniel J. Vanselow
- Mee Siing NguMNMee Siing Ngu
- Carolyn R. ZainoCZCarolyn R. Zaino
- Spencer R. KatzSKSpencer R. Katz
- Yifu DingYDYifu Ding
- Dula ParkinsonDPDula Parkinson
- Khai Chung AngKAKhai Chung Ang
- Patrick J. La RivierePRPatrick J. La Riviere
- Keith C. ChengKCKeith C. Cheng
Preprint
Abstract
Ideal 3-dimensional imaging of many complex samples, such as biological tissues made up of micro-scale structures extending over millimeter- to centimeter-scale tissue samples and organisms, requires both a wide field-of-view and high resolution. With existing optics and detectors used for micro-CT imaging, sub-micron pixel resolution can only be achieved for fields-of-view of Daphnia magna were imaged on Beamline 8.3.2 of the Advanced Light Source. Volumes of 10,000 × 10,000 × 7,096 isotropic 0.5 μm voxels were reconstructed over a 5.0 × 3.5 mm field-of-view. Measurements in the projection domain confirmed a 1.182 μm measured spatial resolution that is largely Nyquist-limited. This unprecedented combination of field-of-view and resolution dramatically reduces the need for sectional scans and computational stitching for large samples, ultimately offering the means to elucidate change in tissue and cellular morphology in the context of larger whole, intact model organisms and specimens. This development is also anticipated to benefit micro-CT imaging in materials science, microelectronics, agricultural science, and biomedical engineering. Synopsis: A custom wide-field lens and a new-generation megapixel camera enabled microCT scanning over a 3.5 × 5 mm field-of-view at a 1 μm resolution / 0.5 μm pixel size at the Berkeley Lawrence Advanced Light Source using a phantom with micron scale features. This novel combination of resolution and field-of-view will be broadly applicable to any setting in which micron-scale structures need to be characterized comprehensively in 3 dimensions over mm to cm scale.Keywords
All Related Versions
- Published version: Journal of Synchrotron Radiation, 29 (2), 505.
This publication has 33 references indexed in Scilit:
- High-Spatial-Resolution Three-dimensional Imaging of Human Spinal Cord and Column Anatomy with Postmortem X-ray Phase-Contrast Micro-CTRadiology, 2021
- 7 Tesla MRI of the ex vivo human brain at 100 micron resolutionScientific Data, 2019
- Computational 3D histological phenotyping of whole zebrafish by X-ray histotomographyeLife, 2019
- Laser‐preparation of geometrically optimised samples for X‐ray nano‐CTJournal of Microscopy, 2017
- Contrast-Enhanced X-Ray Micro-Computed Tomography as a Versatile Method for Anatomical Studies of Adult ZebrafishZebrafish, 2016
- The Ecoresponsive Genome of Daphnia pulexScience, 2011
- A combination of synchrotron and laboratory X-ray techniques for studying tissue-specific trace level metal distributions in Daphnia magnaJournal of Analytical Atomic Spectrometry, 2008
- Developments in synchrotron x-ray computed microtomography at the National Synchrotron Light Source1999
- Detection of Single Mammalian Cells by High-Resolution Magnetic Resonance ImagingBiophysical Journal, 1999
- Within–and between–population variation for resistance of Daphnia magna to the bacterial endoparasite Pasteuria ramosaProceedings Of The Royal Society B-Biological Sciences, 1998